125 V range |
125 V DC ( testing voltage) , 4.000 M& to 40.00 M& ( measurement range) , Accuracy at first effective measurement range ± 2 % rdg. ± 3 dgt. at 0.100 to 10.00 M& |
250 V range |
250 V DC ( testing voltage) , 4.000 M& to 2000 M& ( measurement range) , Accuracy at first effective measurement range ± 2 % rdg. ± 3 dgt. at 0.200 to 20.00 M& |
500 V range |
500 V DC ( testing voltage) , 4.000 M& to 2000 M& ( measurement range) , Accuracy at first effective measurement range ± 2 % rdg. ± 3 dgt. at 0.200 to 50.00 M& |
1000 V range |
1000 V DC ( testing voltage) , 4.000 M& to 4000 M& ( measurement range) , Accuracy at first effective measurement range ± 2 % rdg. ± 3 dgt. at 0.200 to 999 M& |
Low resistance
( conductivity) |
0 to 400.0 ohm, ± 2 % rdg. ± 8 % dgt. , ( aural warning below: 30 ohm) , Open terminal voltage: 4 V max. |
AC voltage range |
0 to 600 V ( 50 to 60 Hz) ± 2 % rdg. ± 8 dgt. , Input impedance: 170 k-ohm |
Other function |
* Insulation resistance mode |
comparator, memory ( 20 data) , measurement value hold, auto discharge, bar graph display, auto display of measurement value 1 minute after measurement start |
* All measurement mode |
live wire warning, battery indicators, auto power save |
Display |
LCD, max. 4000 dgt., Bar graph 42 segment with backlight |
Power supply |
R6P( AA) manganese batteries × 4 or LR6( AA) alkaline batteries × 4 |
Dimensions, mass |
155 mm( 6.10 in) W × 98 mm( 3.86 in) H × 80 mm( 3.15 in) D, 500 g ( 17.7 oz) |
Accessories |
TEST PROBE 9294 ( 1) , Display cover ( 1) , Suspension band ( 1) |