DC Voltage Range: |
420.0 mV to 1000 V, 5 ranges
Best accuracy: ?0.5 % rdg. ?2 dgt |
AC Voltage Range: |
420.0 mV to 1000 V, 5 ranges
Best accuracy: ?1.2 % rdg. ?3 dgt. , Effective value rectifier |
Frequency Characteristics (AC voltage): |
420.0 mV: 50 to 100 Hz, 4.200 to 1000 V: 50 to 500 Hz |
Resistance Range: |
420.0 ohm to 42.00 M-ohm, 6 ranges
Best accuracy: ?0.7 % rdg. ?2 dgt. |
DC Current Range: |
42.00 ?A to 10.00 A, 6 ranges
Accuracy: ?1.5 % rdg. ?4 dgt. |
AC Current Range: |
42.00 ?A to 10.00 A, 6 ranges
Best accuracy: ?2.5 % rdg. ?5 dgt. , Effective value rectifier |
Frequency Characteristics (AC current): |
50 to 500 Hz |
Frequency Range: |
0.50 Hz to 500.0 kHz, 5 ranges
Input level: 800 mV to 1000 V rms
Accuracy: ?0.02 % rdg. ?1 dgt |
Continuity: |
Possible (50 ohm ?30 ohm) or less |
Diode Check: |
Possible (3.4 V open terminal voltage) |
Other Functions: |
Hold-auto function, Duty factor measurement, Relative function, 11 A/1000 V Bussmann fuse |
Auto Power Save: |
Possible (Cancel possible) |
Range Switching: |
Auto or Manual |
Display: |
LCD, maximum 4200 dgt. (all mode), maximum 19999 dgt. (Frequency)
Bar graph display: Maximum 42 segments |
Sampling Rate: |
2.5 times/sec (all mode), 5 times/sec (Frequency), 25 times/sec (Bar graph) |
Power Supply: |
R03(AAA) ? 2 dry batteries (Continuous use 100 hours) |
Dimensions: |
76 mm(2.99 in)W ? 167 mm(6.57 in)H ? 33 mm(1.30 in)D |
Weight: |
260 g (9.2 oz) |
Accessories: |
9207-10 TEST LEAD (1), 3257-51 only: Holster (1) |