Measurement Items: |
Low resistance, AC 4-terminal method |
Generator Section: |
Current generator principle: PWM constant current control
Current setting range: 3.0 A to 31.0 A (0.1 A resolution), into 0.1 Ohm load
Maximum output power: 130 VA*1 (at output terminals)
*1 Subject to derating according to ambient temperature (80 % at 40 °C)
Frequency: 50 Hz or 60 Hz sine wave
Soft start function: Apply current only after checking load connection |
Monitor Section: |
Resistance measurement: 0 to 1.800 Ohm (0.001 Ohm resolution)
Accuracy: ±2 % rdg. ±4 dgt. (after zero-adjust)
Current monitoring range: 0 to 35.0 A AC (0.1 A resolution)
Monitoring cycle: 2 times/second |
Other Functions: |
Timer setting: Counts down time after start until preset time, or shows elapsed time after start
Setting range: 0.5 to 999 second
Comparator: Pass/Fail evaluation using preset upper/lower limit, I/O output
Memory function: max. 20 settings (with save/load) |
Display |
Fluorescent tube digital display |
Power Supply: |
100 to 120 V, 200 to 240 V AC (auto-switching), 50/60 Hz |
Dimensions: |
320 mm(12.60 in)W × 90 mm(3.54 in)H × 263 mm(10.35 in)D |
Accessories: |
Power cord(1), Spare fuse(1), Shorting bar(2) |