Measurement Mode: |
Earth leakage current / Leakage current between enclosure and earth / Leakage current between enclosure and enclosure / Leakage current between enclosure and line / Patient leakage current I / Patient leakage current II / Patient leakage current III / Patient auxiliary current |
Target Current: |
DC / AC / AC+DC (25mA max), AC peak (75mA max) |
Measurement Range: |
DC / AC / AC+DC mode ; 50 ? A / 500 ?A / 5 mA / 25 mA
AC peak mode ; 500 ?A / 1 mA / 10 mA / 75 mA |
Measurement System: |
Indication of a current value calculated based the measured drop in voltage caused by simulated resistance of the human body
Measurement of true effective value
The measurement section ; chassis-grounded and floating |
Accuracy: |
DC / AC / AC+DC mode: ? (2.0 % rdg. + 6 dgt.)
AC peak mode: ? (2.0 % rdg. + 2 dgt.)
DC mode: ? (2.0 % rdg. + 3 dgt.) |
Input Resistance: |
1 M-ohm ? 1 % (Excluding voltmeter section, simulated resistance of the human body) |
Network: |
For medical electrical equipment / For IEC 60990 / For JIS / For UL / General-purpose 1 / General-purpose 2 |
Functions: |
110 % voltage application function / Wiring check function / Automatic measurement function / Application line selection function / Ground fault prevention function / Setting of single-fault condition / Switching power supply polarity / Setting of measuring time / Measurement delay / Maximum value hold / Allowable value judgement / Data save / Clock / Data back up / etc |
Interface: |
EXT I/O , RS-232C , GP-IB |
Power Supply: |
100, 120, 220, 240 V AC (default setting) (50/60 Hz, Rated power ; 30VA) |
Dimensions: |
320 mm(12.6 in)W ? 110 mm(4.33)H ? 263 mm(10.35)D |
Accessories: |
9170 TEST LEAD(2), 9195 ENCLOSURE PROBE(1), 9399 CARRYING CASE(1), Alligator clip(3)(2 red, 1 black), AC Power cord(2), Spare fuse(2) |