Special Characteristics: | Thermocouple input (K, E, J, T) |
Measurement Range: | 2 temperature channels, -100.0 to 1000 ?C (T: 400 ?C) (0.1?C resolution) |
Sensor Type: | Thermocouple (K, E, J, T) |
Main Unit Measurement Accuracy: | ?0.1%f.s. ?2.0 ?C (incl. reference contact compensation) (When connecting an external reference contact compensation sensor between No.2 pin and No.3 pin on Channel 1) |
Reference Contact Compensation Accuracy: | ?2.0 ?C (Reference Contact Compensation Range: 0 to 50 ?C ) (Including Pt 1000 Tolerance: ?0.2%rdg. ?0.15 ?C) |
Temperature Coefficient: | Add ??0.05 ?C per ?C ? to the ?Main Unit Accuracy? (0 to 18 ?C, 28 to 50 ?C) |
Range of Guaranteed Accuracy: | 23 ?C ?5 ?C (Temperature on Main Unit) |