Measurement items |
Insulation resistance (DC voltage application method) |
Output voltage |
25 V to 1000 V (user-configurable), 1 V resolution |
Sampling |
FAST: 30 ms/time, SLOW: 500 ms/time (switch) |
Memory function |
Saved items: rated measurement voltage, comparator upper
limit/lower limit values, test mode, beep sound to distinguish
the result, test time, response time, resistance range,
measurement speed
Memory capacity: up to 10 items (can be saved/loaded) |
Test modes |
Continuous mode, PASS STOP mode, FAIL STOP mode, force
quit assessment mode (switchable) |
Check function |
Contact check function (ON/OFF)
Short-circuit check function (ON/OFF) |
Operating temperature /
humidity range |
0°C to 40°C, 80% rh or lower (non-condensing) |
Operating temperature /
humidity range |
0°C to 40°C, 80% rh or lower (non-condensing) |
Storage temperature/
humidity range |
-10°C to 50°C, 80% rh or lower (non-condensing) |
Guaranteed accuracy
temperature/humidity range |
23°C ±5°C, 80% rh or lower (non-condensing) |
Usage location |
Indoor use, pollution degree 2, up to a height of 2000 m |
Rated power supply voltage |
100 V AC to 240 V ±10% |
Rated power supply
frequency |
50/60 Hz |
Max. rated power |
25 VA |
Withstand voltage |
1.62 kV AC (sensed current 10 mA) 1 min
Between power supply LN (together) and protective grounding
terminal |
Excessive input protection |
1100 V DC (positive polarity only) |
Dimensions |
215 (8.46 in) W x 80 (3.15 in) H x 166 (6.54 in) D mm
(excluding protruding parts) |
Mass |
1.1 kg ± 0.1 kg (38.8 oz ± 3.5 oz) |
Compliant Standards |
Safety EN61010 EMC EN61326 Class A
EN61000-3-2, EN61000-3-3 |
Accessories |
Instruction manual, power cord,
EXT. I/O connector (male), connector cover (one of each) |
Assessment |
UPPER_FAIL, PASS, LOWER_FAIL, UL_FAIL
UPPER_FAIL: Measured value upper limit value
PASS: Upper limit value measured value lower limit value
LOWER_FAIL: Measured value lower limit value
UL_FAIL: Unable to assess |
Assessment Process |
Beep sound, PASS/U.FAIL/L.FAIL light up on LED display
When UL_FAIL, U.FAIL/L.FAIL light up simultaneously
EXT. I/O output, assessment result can be obtained via
RS-232C |
Definition of test duration |
Test duration = Response time + Measurement time |
Function |
Set the time from voltage application until pass/fail assessment |
Configuration range |
0.045 s to 999.999 s (0.001 s resolution) |
Function |
The response time is the time during which comparator
assessment operation is prohibited from the start of the test until
the set response time has elapsed.
No measurement values are displayed during the response time.
The response time is included in the test time. |
Configuration range |
0.005 s to 999.999 s (0.001 s resolution) |