General Specifications |
Controllable devices |
LR8510, LR8511, LR8512, LR8513, LR8514, LR8515 |
No. of controllable devices |
Max. 7 units (up to 105 channels when used with the LR8510 or LR8511) |
Control and communications (Between instrument and units |
Bluetooth® 2.1 + EDR (between Wireless Logging Station and logging modules); communication range: 30 m (line of sight), SSP security |
Internal buffer memory |
8 MWords volatile RAM (SDRAM) |
Clock functions |
Auto calendar, clock accuracy: ±3 s/day (@23°C, 73.4°F) |
Timebase accuracy |
±0.2 s/day while measuring (@23°C, 73.4°F) |
Backup battery life |
At least five years for clock and settings (@23°C, 73.4°F) |
Operating temp. & humidity |
-10 to 50°C (14 to 122°F), 30 to 80%RH or less (non-condensating) |
Storage temp. & humidity |
-20 to 60°C (-4 to 140°F), 80% RH or less (noncondensating) |
Applicable standards |
Safety: EN61010 EMC: EN61326 classA, EN61000-3-2, EN61000-3-3 Wireless certification: Japan (type certification) US(FCC) Canada(IC) EU :Incorporates a wireless module that has been certified as compliant with applicable technical standards. : Part 15.247 (Contains FCC ID: QOQWT11IA) : RSS-210 (Contains IC: 5123A-BGTWT11IA) : EN 300 328, EN 301 489-1, EN 301 489-17 |
Vibration endurance |
JIS D 1601:1995 5.3(1), Category 1: Vehicle, Condition: Category A equiv. |
External control terminal |
External trigger input, trigger output, four alarm channel outputs, ground |
Dimensions and Mass |
230mm (9.06in)W × 125mm (4.92in)H × 36mm (1.42in)D, 700 g (24.7oz.) (excluding Battery Pack) |
Accessories |
Instruction manual ×1, Measurement guide ×1, SD Memory Card (2GB) Z4001 × 1, CD-R (data collection software "Logger Utility") ×1, USB cable ×1, AC Adapter Z1008 × 1 |
Data storage media |
SD memory card |
SD standard-compliant × 1, Hioki Z4001 (2 GB), Data format: FAT16, FAT32 |
USB memory |
Series A receptacle |
Communication functions |
LAN Interface |
IEEE802.3 Ethernet 100BASE-TX DHCP, DNS
- Data acquisition and measurement criteria setting with the Logger Utility
- Setting and measurement by communications commands
- Manual file transfer by FTP server (from the instrument memory or removable storage).
- Auto sending files by FTP client
- Remote control by HTTP server
- E-Mailing
|
USB Interface |
USB2.0 compliant High Speed, Series-mini B receptacle
- Data acquisition, condition settings used with the Logger Utility software (supplied as standard)
- Configure the unit and measure using communication commands
- Transfer data from the SD memory card to a PC via USB drive mode (data transfer not possible from USB memory sticks)
|
Display section |
Display |
5.7 inch TFT color liquid crystal display (640 × 480 pixel), horizontal 16 division, vertical 10 division, selectable between English and Japanese displays, back light saver available |
LCD Brightness |
Selectable from 100, 70, 40, or 25 % |
Power supplies |
AC adapter |
Using the AC Adapter Z1008 (supplied as standard, 100 to 240 VAC, 50/60 Hz), Power consumption: 8 VA (with battery pack removed and maximum brightness) |
Battery |
Using the Battery Pack Z1007 (Li-ion 7.2V 2170mAh) (optional accessory, AC adapter has priority when used in combination with battery pack), continuous operation time: 3 hours (at 23 °C, LCD brightness 25 %) Fast recharging time: 7 hours (the AC Adapter or a 10 to 28 V DC external power supply can be connected while the Battery Pack Z1007 is installed.) |
External power |
10 to 28 VDC (Please contact your HIOKI distributor for connection cord)
15 VA (when battery is charged, and w/LCD max. blightness) |
Trigger functions |
Trigger mode, timing |
Modes : Single / Repeat, Timing : Start / Stop / Start & Stop, Logical sum (OR) and product (AND) of each trigger source, selectable for each channel |
Analog signal source |
Up to 105 channels, depending on how many Wireless Voltage/ Temp Units LR8510 and Wireless Universal Units LR8511 are connected (U1-1 to U7-15). [Level trigger] Triggers when rising or falling through preset level [Window] Triggers when entering or exiting range defined by preset upper and lower limit values [Pattern trigger] Applies the trigger when a pattern defined in terms of 1, 0, ×, and values is matched (Setting only available when using logic measuremeent with the LR8512) |
Interval trigge |
Set year, month, date, hour, minute and second (triggers when specified measurement interval is passed) |
Trigger output |
Open-drain output, Trigger output terminal: Push-button type terminal block (5 V voltage output, active low, pulse width: at least 100 ms) Output response time: Recording interval + 3 sec. or less (with 1 measurement unit, good communications) Recording interval + 5 sec. or less (with 7 measurement units, good communications) |
Alarm output |
Number of channels |
4 channels, non-isolated (common ground with chassis) |
Alarm source |
Analog input: Up to 105 channels, depending on how many Wireless Voltage/ Temp Units LR8510 and Wireless Universal Units LR8511 are connected (U1-1 to U7-15). When thermocouple burn-out detection is enabled, when the Wireless Voltage/ Temp Unit LR8510 or Wireless Universal Unit LR8511 battery is low, or when a communications error occurs |
Alarm type |
Level, window, output latch/ no latch, cancel alarm while measuring |
Alarm sound |
Buzzer, ON/OFF possible |
Alarm output |
Open drain output (with 5 V pull-up, active low), output response time: Recording interval + 3 sec. or less (with 1 measurement unit, good communications) Recording interval + 5 sec. or less (with 7 measurement units, good communications) |
Output sink current |
200 mA at 5 V to 30 VDC |
Measurement settings |
Recording intervals (sampling period) |
*1 , *2 100 ms, *2 200 ms, 500 ms, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s, 1 min, 2 min, 5 min, 10 min, 20 min, 30 min, 1 h (16 selections) All input channels are scanned at high speed during every recording interval *1 Setting not available when the thermocouple burnout detection setting is on. *2 The data update rate of the LR8513 and LR8514 is 500ms |
Recording length (time span) |
Enable continuous recording ON (records until the Stop key is pressed), or continuous recording OFF (enable a specified time span) |
Repeat measurement recording |
Set Off or On.
When On, measurement repeats at the set recording interval. |
Display |
Time axis |
200 ms to 1day/divisions |
Voltage axis |
Select by position (magnification can be×100 to ×1/2, 0 Position : Set between -50 to 150%) or upper/ lower limits |
Waveform scrolling |
Time-axis scrolling is available by left/right arrow keys while measuring and when measurement stops (waveform drawing period). |
Jump function |
Selects the displayed span of the waveform |
Monitor function |
Confirm instantaneous values and waveforms without recording data |
Unit battery life remaining display |
Displays the remaining battery life for wirelessly connected units as 1 of 3 levels. |
Signal strength display |
Displays the signal strength for wirelessly connected units as 1 of 3 levels |
Data saving |
Save destination |
Select a SD memory card or USB memory (use only SD memory cards sold by Hioki). |
Storage operation |
Auto: Save waveform data or time divided calculation results in real time
Manual: Push the save key (operation select: item choose/ directly save) |
Real-time saving |
Possible: Waveforms are saved approximately every one minute as binary or text data to the SD memory card or the USB memory (if sampling rate is slower than 1 minute, waveforms are saved at each interval) To the PC: Waveforms are saved to the HDD in the PC via LAN or USB communication when used with the Logger Utility Software. Data can be saved in real time to the SD memory card or USB memory at the same time. |
Split save |
Simple divide: Save waveform data at pre-set times into separate files from the time measurement starts. On schedule: Designate a reference time within 24 hours and save data into separate files at every set time interval starting from the reference time |
Overwriting save |
Endless loop saving: New file overwrites the oldest file when the SD memory card or USB memory capacity runs short |
Remove external media |
Storage media may be removed during real-time save after message confirmation. Upon inserting the storage media again, data saved in internal memory during that time will be saved as a separate file in the media. |
Data protection |
If a power outage occurs or the battery runs out during real-time saving, power is cut off after the file is closed (protection becomes possible approximately 10 min. or more after the instrument is turned on) |
Save types |
Setting condition, waveform data (binary or text style), calculation of numerical value, screen data (compressed BMP), reservation settings |
Reloading data |
Stored binary data can be recalled by the logging station in 8 MB quantities |
Calculation functions |
Numerical value calculations |
Six calculations are available at the same time Average value, peak value, maximum value, time to maximum value, minimum value, time to minimum value |
Data range of calculation |
During measurement or after stopping: Applies calculations to all data in internal buffer memory, or to the time-span specified by A/B cursors. Interval calculation: Calculate values at pre-determined 1 sec to 1 day intervals and display the latest value |
Calculation value save |
Possible: After measuring the last calculated value is automatically saved to the SD memory card or USB memory as a text file Timed save: Save calculated data at pre-determined 1 sec to 1 day intervals as text data to the SD memory card or USB memory in real time |
Waveform calculations |
Calculate sum, difference, product, and quotient between channels, with calculated results displayed as channels W1 to W30 (valid only while measuring, saved in real time with a channel's waveform data.) |
Other functions |
Event marking |
Search: Move to the event number entered and display the waveforms appearing before and after event Number of events: Maximum 1000 per measurement |
A-B cursor |
Measurement: Time difference between A/B cursors, measured value difference, cursor measured value, time Types: Select trace, vertical, or horizontal |
Scaling |
Convert and display the measurement value of each channel as a scaled value |
Rate adjustment function |
Scaling can be set for a channel so that its value is the same as that for UNIT1-CH1 |
Comment entry |
Enter a title or a comment for each channel |
Others |
Start backup, save 5 types setting conditions into main unit, auto set up, start/stop key lock, key-lock, beep sound, schedule, Quick Set function |