Allowable input voltage |
Up to 5 V |
Measured information |
Impedance, voltage, temperature |
Impedance measurement |
Parameters: R, X, Z, θ, Frequency: 0.1 Hz to 1050 Hz, Measurement ranges: 3.0000 mΩ, 10.0000 mΩ, 100.000 mΩ
Testing current: 3 mΩ range: 1.5 Arms, 10 mΩ range: 500 mArms, 100 mΩ range: 50 mArms |
Voltage measurement |
Measurement range: 5.00000 V (single range), Measurement time: 0.1 s (Fast) to 1.0 s (Slow) |
Temperature measurement Range |
-10.0 °C to 60.0 °C, Measurement time: 2.3 s |
Basic accuracy |
Z: ±0.4% rdg. ±(8 to 60) dgt. , V: ±0.0035% rdg. ±5 dgt. , Temp.: ±0.5 °C (at 10.0 to 40.0 °C) |
Functions |
Comparator, self-calibration, sample delay, average, contact check, measurement current error, and other |
Interfaces |
RS-232C/USB (virtual COM port) * Cannot be used simultaneously
EXT. I/O (NPN/PNP can be switched) |
Power supply |
100 to 240 V AC, 50/60 Hz, 80 VA max |
Dimensions and mass |
330 mm (12.99 in) W × 80 mm (3.15 in) H × 293 mm (11.54 in) D, 3.7 kg (130.5 oz) |
Accessories |
Power cord ×1, Instruction manual ×1, Zero-adjustment board ×1, USB cable (A-B type) ×1, CD-R (communication instruction manual, PC application software, USB driver) ×1 |