9102S |
9100S |
Range |
–10°C to 122°C (14°F to 252°F) at 23°C ambient |
35°C to 375°C (95°F to 707°F) |
Accuracy |
±0.25°C |
±0.25°C at 50°C; ±0.25°C at 100°C;
±0.5°C at 375°C |
Stability |
±0.05°C at 0°C
±0.1°C at 122°C |
±0.07C at 50°C; ±0.1°C at 100°C;
±0.3°C at 375°C |
Well-to-Well Uniformity |
±0.2°C with sensors of similar size at equal depths within wells |
Heating Times |
ambient to 100°C: 10 minutes |
35° to 375°C: 9.5 minutes |
Stabilization |
7 minutes |
5 minutes |
Cooling Times |
ambient to 0°C: 10 minutes |
375°C to 100°C: 14 minutes |
Well Depth |
4 inches (102 mm) |
4 inches (102 mm); 1/16" (1.6 mm) hole is 3.5" (89 mm) deep |
Removable Inserts |
Available in sizes from 1/16" (1.6 mm) to 7/16" (11.1 mm)
[1/4" (6.4 mm) and 3/16" (4.8 mm) included] |
N/A |
Computer Interface |
RS-232 included with free Interface-it software |
Power |
94-234 VAC (±10%), 50/60 Hz, 60 W; or 12 VDC |
115 VAC (±10%), 1.5 A or 230 VAC (±10%), 0.8 A, specify, 50/60 Hz, 175 W |
Size |
3.9" H x 5.5" W x 6.9" D (99 x 140 x 175 mm) |
2.25" H x 4.9" W x 5.9" D (57 x 125 x 150 mm) |
Weight |
4 lb. (1.8 kg) |
2 lb. 3 oz. (1 kg) |
NIST-Traceable Calibration |
Data at –10°C, 0°C, 25°C, 75°C, 55°C, 100°C, and 122°C |
Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 375°C |