SecuTest III+
Test Instrument per BGV A3 (previously VBG 4), MPG (medial product law) and for routine and type tests
Universal tester for testing the electrical safety of portable electrical equipment used in commercial and medical applications, for post-manufacturing tests, testing after repair and for periodic testing.
The SECUTEST SIII+. . .test instruments can be configured for international utilization. The test socket, user interface language and the desired test regulation can all be configured to this end.
Universal test instrument for testing the electrical safety of:
- after repair and for periodic testing of electrical devices in accordance with DIN VDE 0701-0702:2008
- after repair and for periodic testing as well as for technical safety checks of electrical medical devices per IEC 62353 (VDE 0751-1) and/or DIN EN 60601 on the basis of the MPG1)
- in the production of:
- electrical measuring, control and laboratory devices per DIN EN 61010
- data processing systems per DIN EN 60950
- electrical medical devices per DIN EN 60601
- household appliances per DIN EN 60335
Applications
Testing for the Electrical Safety of Electrical Equipment in Accordance with BGV A3
The test instrument can be utilized for quick and safe testing of repaired or modified electrical devices and as well as for periodic testing in accordance with DIN VDE 0701-0702:2008.
- Protective conductor resistance
- Insulation resistance
- Protective conductor current for SC1 devices
- Contact current for SC2 devices
- Absence of voltage at exposed conductive parts (= contact current)
Measuring methods for leakage current measurements:
- Direct measurement
- Equivalent leakage current
- Residual current
Testing for the Electrical Safety of Electrical Medical Devices in Accordance with the German Medical Product Law (MPG) and the associated Operator’s Regulations
The test instrument with feature KA01 is used for quick and safe testing and measurement of repaired or modified electrical medi- cal devices or their components (e.g. patient ports) in accordance with IEC 62353/DIN EN 62353 (VDE 0751) and EN 60601, see also SECULIFEST oder SECULIFE ST HV.
Observance of technical safety requirements allows the user of the test instrument to operate electrical medical devices in a hazard-free fashion. The safety of the patient is also assured through the use of tested electrical medical devices.
The following are measured in accordance with IEC 62353 (VDE 0751-1) regulations:
- Protective conductor resistance
- Insulation resistance
- Equivalent device leakage current
- Equivalent patient leakage current
- Device leakage current
- Patient leakage current
(AC/DC portions are measured separately)
Measuring methods for leakage current measurements:
- Direct measurement
- Equivalent leakage current
- Residual current
A software upgrade (optional) allows for measurement in accordance with EN 60601 regulations, see features on page 6 (with the following single-fault conditions: voltage at application part, interrupted neutral and interrupted protective conductor, with automatic polarity reversal L-N)
- Protective conductor resistance
- Insulation resistance
- L and N connected to protective conductor
- Application parts connected to protective conductor
- Earth leakage current, housing leakage current, patient leakage current, patient auxiliary current
The following additional test conditions can be selected:
- Interrupted operational earth and equipotential bonding
- Housing to ground, application parts to ground
Function Test with Power Analysis (also suitable for high power devices under test up to 16 A)
- Line voltage
- Residual current
- Power consumption
- Active and apparent power
- Power factor
- Electrical energy
- On-time
Multimeter Functions
Extensive multimeter functions including temperature measure- ment expand measuring options for the user in a sensible fashion. The following individual measurements can be performed:
- Direct and alternating voltage (momentary and min/max values)
- Resistance
- Voltage against PE, e.g. phase detection
- Current and protective conductor resistance with clip-on meter (accessory)
- Temperature with Pt100 or Pt1000 (accessory)
High-Voltage Test with Direct Voltage (SECUTEST SIII+ H)
The mains plug of the device under test (safety class I and II devices) is connected to the test socket at the test instrument.
The test instrument monitors the mains connection. Incorrect or dangerous mains connection is indicated, and measurement is disabled in the event of danger.
Use of the test instrument for high-voltage testing is trouble-free because DIN VDE 0104 does not apply. The high-voltage test is performed with direct voltage. In order to comply with require- ments for alternating voltage, testing is performed with 1.5-fold direct voltage. This multiplying factor is applied automatically during testing.
This DC high-voltage test complies with EN 60601 3rd edition/ EN 50106 (VDE 0700 part 500), as well as with other standards.
Report Functions
All values required for electrical device approval reports or device log books (e.g. for ZVEH) can be measured with the test instru- ment.
All measured data can be documented and archived with the measurement and test report, which can be saved to memory and printed out from a PC.
The measurement and test report substantiates regular maintenance and testing for users of electrical devices.
The SECUTEST SI+ module (accessory equipment), a memory with integrated interface and keypad which can be mounted inside the lid of the test instrument, expands the applications range of the test instrument.
Sample displays, menu-driven operation:
Test Sequence Setup |
Visual Inspection |
|
|
Protective Conductor Test |
Leakage Current Measurement |
|
|
Sample displays, online help:
Online Help Texts |
Technical Data |
|
|
Schematic Diagrams |
Internal Circuit Diagrams |
|
|
Sample reports with measurement results:
Test Sequence Results |
Function Test Results |
|
|
Features
The SecuTest III+ has unique capabilities for testing the electrical safety of:
- Electrical Medical Devices according to IEC 60601, UL 60601, UL 544, UL 187, and NFPA 99
- Recurring Testing of ME Equipment according to IEC 62353
- Electrical Measuring, Control and Laboratory Devices according to IEC 61010 & UL 61010)
- Safety of Household Appliances according to IEC 60335
- Data Processing Equipment according to IEC 60950 and UL 1950
- High Voltage Testing as per IEC 60950, IEC 61010, IEC 60335 and IEC 60601
Technical Safety Testing for
- Electrical equipment in accordance with DIN VDE 0701, part 1 edition 2000-9, part 200 and part 260 edition 1990
- Information technology devices and equipment in accordance with DIN VDE 0701, part 240 and DIN EN 60950
- Periodic testing per DIN VDE 0702 (BGV A3)
- Electrical medical devices in accordance with DIN VDE 0751 and EN 60601 (with addendum)
- Electrical equipment for measurement, control and laboratory use per EN 61010
- Electrical household appliances per EN 60335
- Electrical equipment in accordance with British standards
Expanded Functions Including:
- Function test with power analysis
- Temperature measurement
- Current measurement (with optional current clamp)
- Voltage and resistance measurement
- Report generating with printer module (PSI module), SecuStore memory adapter or PC software
- Complete measuring system with automated sequences
- Can be controlled with PC software (remote module for PS3, SE-Q.base)
- Partially programmable test sequences (optional database required)
- User interface in the following languages: English, French, German, Italian, Spanish, Czech, Dutch
- Parallel test sockets for devices under test without mains plug
- DKD calibration certificate (optional)
Options:
- Mains connection for England, France, Germany, Italy, Switzerland, Denmark, USA, China, Australia and adapter set
- High-voltage test
- 25 AAC test current for protective conductor measurements
- DBmed database, operation via modem, remote control, direct printing
- Configuration in accordance with customer specifications
- Measurements in accordance with EN 60601, calibration certificate
- 10 test sockets for application parts
- 2 test sockets for equipotential bonding conductor / operational earth electrode
In order to evaluate electrical safety, tests are performed for protective conductor connections, insulation resistance and leakage current (differential current, equivalent leakage current, device leakage current, patient leakage current, contact current).
Technical safety tests for, amongst others:
- Electrical equipment per DIN VDE 0701-0702, 2008-6 edition
- Electrical medical devices per DIN VDE 0751 / IEC 62353
and EN / GOST R (ГОСТ Р) 60601 (optional)
- Electrical equipment for measurement, control and laboratory use per EN 61010
- Electrical household appliances per EN 60335
- Data processing devices and equipment per EN 60950
Expanded functionality
- Automatic adaptation to either 110 V / 60 Hz or 230 V / 50 Hz mains power
- Automatic recognition of safety class (I, II) (with german and french outlet only) and mains connection errors
- Menu-driven test sequences: fully automated or manual
- Safety for the user thanks to integrated personal protection - suitable for persons with basic electro-technical instruction
- Convenient memory and report generating functions (optional)
- Data interface for PC, printer and barcodes
- Report generation with printer module (PSI module), SecuStore memory adapter or PC software
- User interface in various languages: English, French, German, Italian, Spanish, Czech, Dutch
- Parallel test sockets for devices under test without mains plug
- Patient leakage current: AC and DC components measured separately
- Equivalent leakage current measurement with approx. 230 V test voltage
- Function test with power analysis
- Temperature measurement
- Current measurement (with optional current clamp meter)
- Voltage and resistance measurement
- Optional: High-voltage test up to 6 kV DC
- Dimensions: 292 x 130 x 243 mm, weight: 4.0 e.q. 4.5 kg
Test instruments for executing automated test sequences in accordance with standards IEC 62353, DIN VDE 0701 / 0702 / 0751:
- Including automatic detection and indication of whether or not the probe is connected to PE
- Including direct read-out of measurement data via the RS 232 interface (direct printing)
- Including calibration certificate in accordance with DKD