Specifications |
Embedded Microcontroller |
Custom one-chip LSI device |
Display Type |
LCD with green backlight |
Display Size |
2.05 x 1.5 in. (52 x 38mm) |
Parameters Measured |
pH, ORP, conductivity, TDS content, DO content, salt content |
Measurement Range |
For pH: 0 to 14
For ORP: -1999 mV to +1999 mV
For conductivity: up to 200 mS (microSiemens)
For TDS: up to 132,000 ppm
For DO: up to 20 mg/L
For salt: up to 12% by weight |
Measurement Accuracy |
For pH: ± (0.02 pH units + 2 digits)
For ORP: ± (0.5% of reading + 2 digits)
For conductivity: ± 2% of full-scale reading + 1 digit
For TDS: ± 2% of full-scale reading + 1 digit
For temperature: ± 0.8°C/± 1.5°F
For DO: ± 0.4 mg/L in solution
For oxygen: 0.1% for oxygen in air
For salt: ±0.5% of value |
Measurement Resolution |
For pH: 0.01 unit
For ORP: 1 mV
For conductivity: 0.1 uS @ 200 uS full-scale (f.s.), 0.001 mS @2 mS f.s., 0.01 mS @ 20 mS f.s., 0.1mS @200 mS f.s.
For TDS: 0.1 ppm @200 ppm f.s., 1 ppm @2,000 ppm f.s., 10ppm @ 20,000 ppm f.s
100 ppm @ 100,000 ppm f.s.
For temperature: 0.1°F/0.1°C
For DO: 0.1 mg/L in solution
For oxygen: 0.1% in air
For salt: 0.01% of reading |
Temperature Compensation |
Automatic between 32°and 150°F (0°and 65°C) when using optional ATC temperature probe
Manually settable between 32°and 212°F (0°and 100°C)without ATC probe |
Storable/Recallable Readings |
Maximum, minimum |
Data Logging Sampling Time |
1 second to 9 hours |
SD Card Capacity |
1 GB to 16 GB |
Settable Parameters |
Date, time, auto power off, beep sound, sampling time, decimal point or comma decimal division, temperature unit, compensation temperature. Also salt and altitude compensation (DO mode only) |
Operating Temperature |
32°to 122°F (0°to 50°C) |
Operating Relative Humidity |
0 to 85% |
Power Supply |
Six Alkaline “AA” batteries or optional 9-VDC AC adapter |
Power Consumption |
14 mADC (normal operation, with backlight off and SD card not saving data)
37 mADC with backlight on and card saving data
49 mADC with backlight on and card saving data |