General Specifications |
Robust instrument, designed for tough environmental conditions, waterproof (rubber protective cover, IP56). |
General |
Display |
1999 digit LC-display with special symbols, digit height 25 mm, fluorescent backlight |
User interface |
Instant measurement through TURN and START - one button concept . The only operating elements are rotary switch and START button. |
Temperature ranges |
Working temp. |
-10° C ... +50° C ( +14° F ...+122° F ) |
Operating temp. |
0° C ... +35° C ( +32° F ... +95° F ) |
Storage temp. |
-20° C ... +60° C ( -4° F ...+140° F ) |
Reference temp. |
+23° C ±. 2° C ( +73° F ± 4° F ) |
Temp. coefficient |
± 0,1 % of reading / K |
Intrinsic error |
Refers to the reference temperature range and is guaranteed for 3 years. |
Operating error |
Refers to the operating temperature range and is guaranteed for 3 years. |
Climatic class |
C1 (IEC 654-1), -5° C...+45°C, 5%...95% RH |
Protective type |
IP56 according to EN 60529 |
Safety |
Protection by double and/or reinforced insulation. max 50V to earth |
EMC (Emission) |
IEC 61326-1:1997 Class B CISPR16 (CISPR22), CISPR16-1 |
EMC (Immunity) |
IEC61326-1:1997 IEC 61000-4-2 8 kV (air) perf. criteria B, IEC 61000-4-3 3 V/m perf. criteria A, IEC 61000-4-4 0,5 kV perf. criteria B IEC 61000-4-5 1 kV perf. criteria B IEC 61000-4-6 3 V perf. criteria A IEC 61000-4-8 Level 4, 30A/m perf. criteria A |
Quality system |
developed, designed and manufactured according to DIN ISO 9001 |
External voltage |
Max. Uext = 24 V (AC + DC), measurement inhibited for higher values |
Uext rejection |
>120dB (162/3, 50, 60, 400Hz) |
Measuring time |
typical 6 sec |
Max. overload |
250 Vrms |
Auxiliary power |
6 x 1,5 V mignon cells alkali-manganese (type AA LR6) |
Battery life span |
typical > 3000 measurements |
Dimensions |
240 x 180 x 110 mm |
Weight |
1,1 kg (including batteries) |