|
5606 |
5607 |
Temperature range |
-200 °C to 160 °C 0 °C to 450 °C |
Nominal resistance at 0.01 °C |
100 O ± 0.1 O |
Temperature coefficient |
0.00385 O/O/°C |
Accuracy [1] |
± 0.04 °C at 0.01 °C
± 0.06 °C at 160 °C |
± 0.04 °C at 0.01 °C
± 0.07 °C at 450 °C |
Short-term repeatability [2] |
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C |
± 0.020 °C at 0.01 °C
± 0.035 °C at 450 °C |
Drift [3] |
± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C |
± 0.02 °C at 0.01 °C
± 0.04 °C at 450 °C |
Hysteresis |
± 0.015 °C |
Sheath length |
50 mm ± 5 mm
(2 in ± 0.2 in) |
40 mm ± 5 mm
(1.6 in ± 0.2 in) |
Sheath diameter |
3.1 mm ± 0.1 mm (1/8 in ± 0.004 in) |
Sheath material |
316 SST |
Inconel ± 600 |
Transition junction temperature range [4] |
-200 °C to 160 °C |
0 °C to 450 °C |
Transition junction dimensions |
No transition |
35 mm x 7 mm (1.4 in x 0.3 in) |
Sensor length |
30 mm ± 3mm (1.2 in ±0.1 in) |
Sensor location |
3 mm ± 1 mm from tip (1.2 in) |
Minimum insulation resistance |
20 MO at 23 °C |
100 MO at 23 °C |
Minimum immersion length |
Full immersion |
Full immersion |
Maximum immersion depth in liquid medium |
Full immersion |
Use SST or glass tube to protect transition junction |
Maximum immersion depth in dry medium |
Full immersion |
Full immersion |
Response time [5] |
12 seconds typical |
9 seconds typical |
Self heating (in 0 °C bath) |
± 0.003 °C |
Lead-wire cable type |
Enameled copper wire |
SST over-braid |
Lead-wire length |
2.4 meters (8 ft) |
Lead-wire temperature range |
160 °C |
450 °C |
Calibration |
Calibration not included. NVLAP accredited calibration available. See ordering information. |
[1] Includes calibration and 100 hr drift (k = 2).
[2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k = 2).
[3] After 100 hours at max temp, 95 % confidence (k = 2).
[4] Temperatures outside this range will cause irreparable damage.
[5] Per ASTM E 644. |