Technical Data |
Number: |
4 voltage and current (3 phases + neutral) |
Maximum voltage: |
1000 Vrms (6 kV peak) |
Vrms (AC + DC) Measurement range: |
1 ... 1000 V |
Accuracy: |
0.1% of Vnom |
Vpeak Measurement range: |
1 ... 1400 V |
Accuracy: |
5% of Vnom |
Crest factor, voltage Measurement range: |
1.0 ... > 2.8 |
Accuracy: |
?5% |
Arms (AC + DC) Measurement range: |
0 ... 20 kA |
Accuracy: |
?0.5% ? 5 counts |
Apeak Measurement range: |
0 ... 5.5 kA |
Accuracy: |
5% |
Crest factor, A Measurement range: |
1 ... 10 |
Accuracy: |
5% |
Hz 50 Hz nominal Measurement range: |
40 ... 70 Hz |
Accuracy: |
?0.01 Hz |
Vrms (AC+DC) ? Measurement range: |
0.0% ? 100% of Vnom |
Accuracy: |
?0.2% of nominal voltage |
Arms (AC+DC) ? Measurement range: |
0 ... 20 kA |
Accuracy: |
?1% ? 5 counts |
Harmonic (interharmonic) (n) Measurement range: |
DC, 1..50; (Off, 1..49) measured according to IEC 61000-4-7 |
Vrms Measurement range: |
0.0 ? 1000 V |
Accuracy: |
?0.05% of nominal voltage |
Arms Measurement range: |
0.0 ? 4000 mV x clamp scaling |
Accuracy: |
?5% ? 5 counts |
Watts Measurement range: |
Depends on clamp scaling and voltage |
Accuracy: |
?5% ? n x 2% or reading, ? 10 counts |
DC voltage Measurement range: |
0.0 ? 1000 V |
Accuracy: |
?0.2% of nominal voltage |
THD Measurement range: |
0.0 ? 100.0% |
Accuracy: |
?2.5% V and A (? 5% Watt) |
Hz Measurement range: |
0 ? 3500 Hz |
Accuracy: |
? 1 Hz |
Phase angle Measurement range: |
-360? ... +360? |
Accuracy: |
? n ? 1.5? |
Watt, VA, VAR Measurement range: |
1.0 ? 20.00 MVA? |
Accuracy: |
?1% ? counts |
kWh, kVAh, kVARh Measurement range: |
00.00 ?200.0 GVAh? |
Accuracy: |
? 1.5% ? 10 counts |
Power Factor/ Cos ⊕ / DPF Measurement range: |
0?1 |
Flicker |
? 0.03 |
Pst (1 min), Pst, Plt, PF5 Measurement range: |
0.00 ? 20.00 |
Accuracy: |
?5% |
Volts Measurement range: |
0.0 ? 5.0% |
Accuracy: |
?0.5% |
Current Measurement range: |
0.0 ? 20% |
Accuracy: |
? 1% |
Volts Measurement range: |
?6000 V |
Accuracy: |
?2.5% of Vrms |
Minimum detect duration |
5 ?s (200kS/s sampling) |
Arms (AC+DC) Measurement range: |
0.000 ? 20.00 kA? |
Accuracy: |
?1% of meas ? 5 counts |
Inrush duration (selectable) Measurement range: |
7.5 s ... 30 min |
Accuracy: |
?20 ms (Fnom = 50 Hz) |
Sampling: |
5 readings/sec continuous sampling per channel |
Memory: |
1800 min, max and avg points for each reading |
Recording time: |
Up to 450 days |
Zoom: |
Up to 12x horizontal zoom |
Screens & data: |
50, shared memory divided between logging, screens and data sets |
Notes |
? Depending on clamp scaling
? Value is measured over 1 cycle, commencing at a fundamental zero crossing, and refreshed each half-cycle |
The advanced functions Interharmonics, Energy, Transients and Inrush are optional for the Fluke 434 and standard available on the Fluke 435. |
Environmental Specifications |
Operating Temperature: |
0 ?C to +50 ?C |
Safety Specifications |
Safety |
EN61010-1 (2nd edition) pollution degree 2; 1000 V CAT III / 600 V CAT IV ANSI/ISA S82.01 |
Mechanical & General Specifications |
Size |
256 x 169 x 64 mm |
Weight |
1.1 kg |
Rechargeable NiMH pack (installed): |
>7 hours |
Battery charging time: |
4 hours typical |
Shock: |
30 g |
Vibration: |
3 g according to MIL-PRF-28800F Class 2 |
Case |
Rugged, shock proof with integrated protective holster, IP51 (drip and dust proof) |
Warranty |