General specifications |
Maximum voltage
applied to any terminal |
1000 V ac rms or dc |
Storage temperature |
-40 °C to 60 °C (-40 °F to 140 °F) |
Operating temperature |
-20 °C to 55 °C (-4 °F to 131 °F) |
Temperature coefficient |
0.05 x (specified accuracy) per °C for temperatures < 18 °C or > 28 °C (< 64 °F or > 82 °F) |
Relative humidity,
non-condensing |
< °C
0 % to 95 % @ 10 °C to 30 °C (50 °F to 86 °F)
0 % to 75 % @ 30 °C to 40 °C (86 °F to 104 °F)
0 % to 40 % @ 40 °C to 55 °C (104 °F to 131 °F) |
Vibration |
Random, 2 g, 5-500 Hz per MIL-PRF-28800F, Class 2 instrument |
Shock |
1 meter drop per IEC 61010-1 2nd Edition (1 meter drop test, six sides, oak floor) |
Electromagnetic
compatibility |
In an RF field of 3 V/M, accuracy = specified accuracy except in temperature: accuracy =
specified accuracy ± 5 °C (9 °F). (EN 61326-1:1997) |
Safety |
Complies with ANSI/ISA 82.02.01 (61010-1) 2004, CAN/CSA-C22.2 NO. 61010-1-04, and
IEC/EN 61010-1 2nd Edition for measurement CAT III 1000 V and CAT IV 600 V |
Certifications |
CSA per standard CSA/CAN C22.2 No. 61010.1-04; TUV per standard EN 61010 Part 1-1002 |
Batteries |
Four AA batteries (NEDA 15A or IEC LR6) |
Battery life |
Meter use 1000 hours; Insulation test use: Meter can perform at least 1000 insulation tests
with fresh alkaline batteries at room temperature. These are standard tests of 1000 V into
1 MΩ with a duty cycle of 5 seconds on and 25 seconds off |
Size |
5.0 cm H x 10.0 cm W x 20.3 cm L (1.97 in H x 3.94 in W x 8.00 in L) |
Weight |
550 g (1.2 lb) |
IP rating |
IP40 |
Altitude (operating) |
2000 m CAT III 1000 V, CAT IV 600 V; 3000 m CAT II 1000 V, CAT III 600 V |
Storage |
12,000 m |
Over-range capability |
110 % of range except for capacitance which is 1 % |
Compliance to EN 61557 |
IEC61557-1, IEC61557-2 |