1523 |
1524 |
Input Channels |
1 |
2 |
Logging |
25 readings with statistics |
25 readings with statistics
15,000 time and date stamped |
Sample interval (normal): |
1 second |
1 second (simultaneous measurement) |
Typical sample interval (fast mode)*: |
0.3 seconds |
0.3 seconds |
Sensor types |
PRTs, RTDs, Thermistors, and Thermocouples |
Thermocouple types |
C,E,J,K,L,M,N,T,U,B,R,S |
Operating temperature |
-10 °C to 60 °C (Best accuracy 13 °C to 33 °C) |
Power requirements |
3 AA alkaline batteries |
Safety |
EN 61010-1:2001, CAN/CSA C22.2 No. 61010.1-04 |
Size (HxWxL) |
96 mm x 200 mm x 47 mm
(3.75 in x 7.9 in x 1.86 in) |
Weight |
0.65 kg (1.4 lb) |
Environmental conditions for best accuracy: 13 °C to 33 °C |
Millivolt range and accuracy |
-10 mV to 75 mV
± (0.005 % + 5 V) |
Resistance range and accuracy |
0 Ω to 400 Ω ± (0.004 % + 0.002 Ω)
200 Ω to 50 kΩ ± (0.01 % + 0.5 Ω)
50 kΩ to 500 kΩ ± (0.03 %) |
Temperature coefficient, voltage ( -10 °C to 13 °C, + 33 °C to 60 °C) |
Temperature coefficient
( -10 °C to 13 °C , +33 °C to 60 °C):
± (0.001 %/°C + 1 V/°C) |
Temperature Coefficient, Resistance ( -10 °C to 13 °C, + 33 °C to 60 °C) |
0.0008 %/ °C + 0.0004 Ω (0 Ω to 400 Ω)
0.002 %/ °C + 0.1 Ω (0 Ω to 50 kΩ)
0.06 %/ °C + 0.1 Ω (50 kΩ to 500 kΩ) |
Excitation current, resistance |
1 mA (0 Ω to 400 Ω)
10 µA (0 Ω to 50 kΩ)
2 µA (50 kΩ to 500 kΩ) |
*See technical manual for sample interval details by sensor type and number of inputs. |