Power Meter Specifications |
Input Connector |
Interchangeable connector adapter (LC standard, SC, ST and FC optional) |
Detector Type |
InGaAs |
Wavelengths |
850 nm, 1300 nm, 1310 nm, 1550 nm |
Power Measurement Range |
0 dBm to -65 dBm (850 nm) 0 dBm to -70 dBm (all other wavelengths) |
Power Measurement Uncertainty1,2 |
< +/- 5% +/- 32 pW |
Measurement Linearity3 |
< ± 0.1 dB |
Re-calibration Period |
1 year |
1. +/- 100 pW at 850 nm
2. Under the following conditions: Power level 100 μW (-10 dBm), continuous wave (CW) for absolute power at 850 nm and 1310 nm. Divergent beam, NA = 0.20 for 50/125 μm and NA = 0.14 for 9/125 μm. Ambient temperature 23° ± 1°C. SC/UPC connector with ceramic ferrule. After a 5 minute warm-up. Traceable to NIST.
3. -3 dBm to -55 dBm at 850 nm and 1310 nm. Ambient temperature 23° ± 1°C. After a 5 minute warm-up. |
Loss/Length Specifications |
Specification |
CertiFiber Pro Multimode Modules |
CertiFiber Pro Singlemode Modules |
Testing Speeds (excludes referencing times) |
Smart Remote mode: < 3 seconds (2 wavelengths, one direction, auto wavelength detection)
Far End Source mode: ≤ 2 seconds (2 wavelengths, one direction auto wavelength detection)
Loopback mode: ≤ 2 seconds (2 wavelengths, one direction auto wavelength detection) |
Input/Output Connectors |
Interchangeable connector adapter (LC standard, SC, ST and FC optional) |
Launch Condition1,2 |
Encircled Flux compliant to TIA-526-14-B, ISO/IEC 14763-3 and IEC 61280-4-1 |
Fiber Types Tested |
50/125 µm or 62.5/125 µm |
singlemode |
Source Type and Wavelengths |
LED source 850 nm ± 30 nm 1300 nm ± 20 nm |
Fabry-Perot laser diode 1310 nm ± 20 nm 1550 nm ± 30 nm |
Maximum Length Measurement |
12 km |
130 km |
Length Measurement Accuracy |
± 1.5 m plus ± 1 % of length |
± 1.5 m plus ± 1 % of length |
Output Power (Nominal) |
≥ -24 dBm with EF-TRC |
≥ -4 dBm |
Output Power Stability3 |
± 0.05 dB over 8 hours ± 0.03 dB over 15 minutes |
± 0.1 dB over 8 hours ± 0.08 dB over 15 minutes |
1. At the output of the EF-TRC
2. Variations between EF measurement equipment may occur but EF compliance can be expected with a 95% confidence factor
3. Relative to power level after 15 minute warm up – at constant temperature |
Loss/Length Specifications (continued) |
Source Wavelengths |
850 nm, 1300 nm, 1310 nm, 1550 nm |
Power Measurement Range |
0 dBm to -65 dBm (850 nm) 0 dBm to -70 dBm (all other wavelengths) |
Power Measurement Uncertainty1, 2 |
< +/- 5% +/- 32 pW |
Measurement Linearity3 |
< ± 0.1 dB |
Re-calibration Period |
1 year |
1. +/- 100 pW at 850 nm
2. Under the following conditions: Power level 100 μW (-10 dBm), continuous wave (CW) for absolute power at 850 |
Visual Fault Locator (VFL) |
Title |
Requirement |
Output Power |
>-5 dBm
<0 dBm
SMF-28 singlemode fiber
Continuous wave
SC/UPC connector |
Operating Wavelength |
650 nm nominal |
Output Modes |
Continuous wave
Pulsed mode (2 Hz to 3 Hz blink frequency) |
Connector Adapter |
2.5 mm universal |
Laser Safety |
Class II CDRH
Over operating temperature range |
Environmental Specifications |
Title |
Requirement |
Operating Temperature |
-18° C to 45° C |
Storage Temperature |
-30° C to +60° C |
Operating Relative Humidity
% RH without condensation |
0 % to 90 %, 32 °F to 95 °F (0° C to 35° C)
0 % to 70 %, 95 °F to 113 °F (35° C to 45° C) |
Vibration |
Random, 2 g, 5 Hz-500 Hz |
Shock |
1 m drop test with and without module and adapters |
Safety |
CSA C22.2 No. 1010.1: 1992
EN 61010-1 1st Edition + Amendments 1, 2 |
Pollution Degree |
2 |
Altitude |
Operating: 4,000 m; storage: 12,000 m |
EMC |
EN 61326–1 |