Model: SMAN460 Specifications |
Mini-USB port |
For updating to newer versions of firmware |
Display size |
5 inches (diagonal) |
Backlight |
Blue (On for 3 minute unless turned off manually) |
Battery |
6 x AA (Battery life below based on alkaline type) |
Battery life |
350 hours (without vacuum, backlight and wireless) |
Low battery indication |
An indicator is displayed when the battery voltage drops below the operating level |
Auto shut off |
30 minutes of inactivity when APO is activated |
Over range |
"OL" or "-OL" is displayed |
Operating environment |
32°F to 122°F (0°C to 50°C) at <75% RH |
Temperature coefficient |
0.1 x (specified accuracy) per °C (0°C to 18°C, 28°C to 50°C), per 0.6°F (32°F to 64°F, 82°F to 122°F) |
Storage temperature |
-4°F to 140°F (-20°C to 60°C), 0 to 80% RH (with battery removed) |
Weight |
4.03 lbs (1.83 kg) |
Pressure |
Sensor type |
Absolute pressure sensors |
Connector type |
Standard 1/4" and 3/8" NPT male flare fitting |
Range |
29" HgV to 580psig (English), 74 cmHgV to 0 to 4000KPa (Metric), 4.000MPa (Metric), and 40.00bar (Metric) |
Resolution |
0.1 psi/inHg; 1 kPa/cmHg; 0.001MPa; 0.01bar |
Accuracy |
29" HgV to 0" HgV: ±0.2" HgV 74 cmHgV to 0 cmHgV: ±1 cmHgV 0 to 200 Psig: ±1 Psi; 0 to 1378 KPa: ±7 KPa; 0 to 1.378MPa: ±0.007MPa; 0 to 13.78 bar ±0.07 bar 200 to 580 Psig: ±(0.3% of reading+1 Psig); 1378 to 4000KPa: ±(0.3% of reading+7 Kpa); 1.378 to 4.000MPa: ±(0.3% of reading+0.007MPa); 13.78 to 40.00bar: ±(0.3% of reading +0.07bar) |
Maximum overload pressure |
800 psig |
Units |
Psig, kPa, MPa, bar, inHg, and cmHg |
Microns for Vacuum |
Connector type |
Standard 1/4" and 3/8" NPT male flare fitting |
Range |
0 to 9999 microns of mercury |
Resolution |
1 micron (0 to 2000 microns), 250 microns (2001 to 5000 microns), 500 microns (5001 to 8000 microns), 1000 microns (8001 to 9999 microns) |
Accuracy |
±(5% of reading + 5 microns), 50 to 1000 microns |
Maximum overload pressure |
580 psig |
Units |
Microns of mercury |
Temperature |
Sensor type |
Type K thermocouple |
Range |
-95°F to 999.9°F (-70°C to 537.0°C) |
Resolution |
0.1°F/°C |
Accuracy |
±(1.0°F) -95°F to 199.9°F; ±(0.5°C) -70°C to 93°C ±(2.0°F) 200°F to 999.9°F; ±(1.0°C) 93°C to 537.0°C
Note: All accuracies are after a field calibration. |
Wireless |
Wireless range |
1 to 100 feet (30m) |
Initial sync setup range |
1 to 10 feet
Wirelessly transmit to Fieldpiece HVAC Guide® HG3 System Analyzer
Wirelessly receive indoor wet bulb (IDWB) and outdoor dry bulb
(ODDB) from Fieldpiece transmitters |