Features include:
Zero Modes (AZERO/MZERO)
DR58 starts up in automatic zero mode (AZERO) which sets a new zero point every two seconds, indicated by AZERO blinking twice.
Moving closer to the leak source will show the relative increase in concentration, triggering the alarm. Conversely, stopping at or moving away from the leak source will not trigger a response.
Press MANUAL/AUTO >1 second to toggle manual zero mode (MZERO). DR58 does not zero automatically while MZERO is displayed and will continue to alarm and display the relative leak size.
Zero Set
Press ZERO to manually set the currently measured concentration to 0, indicated by AZERO/MZERO blinking twice.
Numerical Leak Size Indicator with Bar Graph
The number is the relative concentration of refrigerant gas (0-99 units). The bar graph shows the same measurement in steps of 10 units for easy monitoring. The higher the number, the higher the concentration of gas being measured and the closer you are to the exact leak location.
Maximum Mode
Press MAX ON/OFF >1 second to toggle max mode. When max mode is on, the highest leak size detected is continually recorded.
The recorded max is cleared when the sensitivity is changed or if manually zeroed.
Sensitivity (H/M/L)
DR58 starts up in high sensitivity (H) which is recommended for most situations. When leaks are difficult to pinpoint due to excess triggering or sensor saturation, press SENSITIVITY to cycle to a lower sensitivity (Medium/Low).