Diode Test |
Test current of 0.3mA maximum, open circuit voltage 1.5V DC typical |
Continuity Check |
Audible signal will sound if the resistance is less than approximately 30 Ω, test current <0.7mA |
Temperature sensor |
Requires type K thermocouple |
Input Impedance |
7.5M Ω (VDC and VAC) |
Display |
4000 count LCD |
Overrange indication |
“OL” is displayed |
Polarity |
Automatic (no indication for positive polarity); Minus (-) sign for negative polarity. |
Measurement Rate |
2 times per second, nominal |
Auto Power Off |
Meter automatically shuts down after 15 minutes of inactivity |
Low Battery Indication |
“BAT” is displayed if battery voltage drops below operating voltage |
Batteries |
Requires two AAA batteries |
Fuses |
mA, μA ranges, 0.5A/250V fast blow 10A range, 10A/250V fast blow |
Operating Temperature |
32°F to 122°F (0°C to 50°C) |
Storage Temperature |
-4°F to 140°F (-20°C to 60°C) |
Relative Humidity |
<70% operating, <80% storage |
Operating Altitude |
7000ft (2000 meters) maximum. |
Weight |
9.17 oz. (260g). |
Size |
4.78” x 2.38” x 1.57” (121.5mm x 60.6mm x 40mm) |
Safety |
For indoor use and in accordance with Overvoltage Category II, Pollution Degree 2. Category II includes local level, appliance, portable equipment, etc., with transient overvoltages less than Overvoltage Category III. |
Input Limits |
V DC, VAC |
600V DC/AC |
Ohms, Capacitance, Diode, Frequency, Continuity |
250V DC/AC |
mA DC/AC |
400mA DC/AC |
10A DC/AC |
10A DC/AC (30sec. max) |