| Model: EA-PSI 9080-10 DT Specifications |
| Product Type |
DC Laboratory Power Supply |
| Housing |
Desktop (table) |
| Voltage (U) |
80.00 |
| Current (I) Ampere |
10.00 |
| Power (P) Watt |
320.00 |
| Resistance (R) |
Yes |
| Characteristic Curve |
U/I/P autoranging |
| Human Interface |
Touch panel & Menu |
| Ext. control - analogue 0-5V/0-10V |
Yes |
| External control - GPIB (IEEE488) |
No |
| External control - USB |
Yes |
| External control - RS232 |
No |
| External control - CAN |
No |
| External control - LAN |
Yes |
| External control - Profibus |
No |
| External control - DeviceNet |
No |
| External control - Modbus TCP |
Yes |
| Function generator |
Yes |
| Special feature - High-Speed |
No |
| Special feature - Sink |
No |
| Special feature - water cooled |
No |
| AC Input |
| Voltage range |
90...264 V AC |
| Connection |
1ph,N,PE |
| Frequency |
45-65 Hz |
| Fusing |
MT 8 A |
| Leak current |
< 3.5 mA |
| Power factor |
≈ 0.99 |
| DC Output |
| Max. output voltage U Max |
81.6 V |
| Max. output current I Max |
10 A |
| Max. output power P Max |
320 W |
| Overvoltage protection range |
0...88 V |
| Overcurrent protection range |
0...11 A |
| Overpower protection range |
0…352 W |
| Output capacitance |
3400 μF |
| Temperature coefficient for set values Δ/K |
Voltage / current: 100 ppm |
| Voltage regulation |
| Adjustment range |
0...81.6 V |
| Accuracy (1 (at 23 ± 5°C) |
< 0.1% U Max |
| Line regulation at ±10% ΔU AC |
< 0.02% U Max |
| Load regulation at 0...100% load |
< 0.05% U Max |
| Rise time 0...100% (@full load) |
Max. 60 ms |
| Settling time after load step |
< 1.5 ms |
| Display: Accuracy |
≤ 0.2% U Max |
| Ripple |
< 8 mV PP
< 0.8 mV RMS |
| Remote sensing compensation |
Max. 5% U Max |
| Output voltage fall time (at no load) after switching DC output off |
Down from 100% to <60 V: less than 10 s |
| Current regulation |
| Adjustment range |
0...10,2 A |
| Accuracy (1 (at 23 ± 5°C) |
< 0.2% I Max |
| Line regulation at ±10% ΔU AC |
< 0.05% I Max |
| Load regulation at 0...100% ΔU OUT |
< 0.15% I Max |
| Ripple |
< 1 mA RMS |
| Display: Accuracy |
≤ 0.2% I Max |
| Power regulation |
| Adjustment range |
0…326.4 W |
| Accuracy (1 (at 23 ± 5°C) |
< 1% P Max |
| Line regulation at ±10% ΔU AC |
< 0.05% P Max |
| Load reg. at 10-90% ΔU OUT * ΔI OUT |
< 0.75% P Max |
| Power regulation |
| Display: Accuracy |
≤ 0.8% P Max |
| Efficiency |
≈ 92% |
| Internal resistance regulation |
| Adjustment range |
0...163.2 Ω |
| Accuracy |
≤ 1% of max. resistance ± 0.3% of maximum current |
| Analog interface |
| Set value inputs |
U, I, P, R |
| Actual value output |
U, I |
| Control signals |
DC on/off, remote control on/off, resistance mode on/off |
| Status signals |
CV, OVP, OCP, OPP, OT, PF, DC on/off |
| Galvanic isolation to the device |
Max. 400 V DC |
| Sample rate of inputs |
500 Hz |
| Insulation |
| Output (DC) to enclosure (PE) |
DC minus: permanent max. ±400 V
DC plus: permanent max. ±400 V + output voltage |
| Input (AC) to output (DC) |
Max. 2500 V, short-term |
| Miscellaneous |
| Cooling |
Temperature controlled fan, side inlet, rear exhaust |
| Ambient temperature |
0..50°C |
| Storage temperature |
-20...70°C |
| Humidity |
< 80%, not condensing |
| Standards |
EN 60950:2006 + A11:2009 + A1:2010 + A12:2011 + AC:2011 + A2:2013
EN 61236-1:2013-07 |
| Overvoltage category |
2 |
| Protection class |
1 |
| Pollution degree |
2 |
| Operational altitude |
< 2000 m |
| Digital interfaces |
| Featured |
1x USB-B for communication, 1x USB-A for functions, 1x LAN for communication |
| Galvanic isolation from device |
Max. 400 V DC |
| Terminals |
| Rear side |
AC input, analog interface, USB-B, Ethernet |
| Front side |
DC output, USB-A, remote sensing |
| Dimensions |
| Enclosure (WxHxD) |
276 x 103 x 355 mm |
| Total (WxHxD) |
308 x max. 195 x min. 391 mm |
| Weight |
≈ 7.5 kg |
| Article number |
06200501 |