Model: EA-PSI 9040-20 DT Specifications |
Product Type |
DC Laboratory Power Supply |
Housing |
Desktop (table) |
Voltage (U) |
40.00 |
Current (I) Ampere |
20.00 |
Power (P) Watt |
320.00 |
Resistance (R) |
Yes |
Characteristic Curve |
U/I/P autoranging |
Human Interface |
Touch panel & Menu |
ext. control - analogue 0-5V/0-10V |
Yes |
external control - GPIB (IEEE488) |
No |
external control - USB |
Yes |
external control - RS232 |
No |
external control - CAN |
No |
external control - LAN |
Yes |
external control - Profibus |
No |
external control - DeviceNet |
No |
external control - Modbus TCP |
Yes |
function generator |
Yes |
special feature - High-Speed |
No |
special feature - Sink |
No |
special feature - water cooled |
No |
AC Input |
Voltage range |
90...264 V AC |
Connection |
1ph,N,PE |
Frequency |
45-65 Hz |
Fusing |
MT 8 A |
Leak current |
< 3.5 mA |
Power factor |
≈ 0.99 |
DC OutputMax. output voltage UMax |
40.8 V |
Max. output current IMax |
20 A |
Max. output power PMax |
320 W |
Overvoltage protection range |
0...44 V |
Overcurrent protection range |
0...22 A |
Overpower protection range |
0...352 W |
Output capacitance |
4760 μF |
Temperature coefficient for set values Δ/K |
Voltage / current: 100 ppm |
Voltage Regulation |
Adjustment range |
0...40.8 V |
Accuracy (1 (at 23 ± 5°C) |
< 0.1% UMax |
Line regulation at ±10% ΔUAC |
< 0.02% UMax |
Load regulation at 0...100% load |
< 0.05% UMax |
Rise time 0...100% (@full load) |
Max. 30 ms |
Settling time after load step |
< 1.5 ms |
Display Resolution |
See section „1.9.5.4. Resolution of the displayed values in the Operating Guide |
Display Accuracy (3 |
≤ 0.2% UMax |
Ripple (2 |
< 8 mVPP
< 0.8 mVRMS |
Remote sensing compensation |
Max. 5% UMax |
Output voltage fall time (at no load)after switching DC output off |
- |
Current Regulation |
Adjustment range |
0...20,4 A |
Accuracy (1 (at 23 ± 5°C) |
< 0.2% IMax |
Line regulation at ±10% ΔUAC |
< 0.05% IMax |
Load regulation at 0...100% ΔUOUT |
< 0.15% IMax |
Ripple (2 |
< 1 mARMS |
Display Resolution |
See section "1.9.5.4. Resolution of the displayed values" in the Operating Guide |
Display Accuracy (3 |
≤ 0.2% IMa |
Power Regulation |
Adjustment range |
0...326.4 W |
Accuracy (1 (at 23 ± 5°C) |
< 1% PMax |
Line regulation at ±10% ΔUAC |
< 0.05% PMax |
Load reg. at 10-90% ΔUOUT * ΔIOUT |
< 0.75% PMa |
Display Resolution |
See section "1.9.5.4. Resolution of the displayed values" in the Operating Guide |
Display Accuracy (2 |
≤ 0.8% PMax |
Efficiency (4 |
≈ 92% |
Internal Resistance Regulation |
Adjustment range |
0...81.6 Ω |
Accuracy (1 |
≤ 1% of max. resistance ± 0.3% of maximum current |
Display Resolution |
See section "1.9.5.4. Resolution of the displayed values" in the Operating Guide |
Analog interface (3 |
Set value inputs |
U, I, P, R |
Actual value output |
U, I |
Control signals |
DC on/off, remote control on/off, resistance mode on/off |
Status signals |
CV, OVP, OCP, OPP, OT, PF, DC on/off |
Galvanic isolation to the device |
Max. 400 VDC |
Sample rate of inputs |
500 Hz |
Insulation |
Output (DC) to enclosure (PE) |
DC minus: permanent max. ±400 V | DC plus: permanent max. ±400 V + output voltage |
Input (AC) to output (DC) |
Max. 2500 V, short-term |
Miscellaneous |
Cooling |
Temperature controlled fan, side inlet, rear exhaust |
Ambient temperature |
0..50°C |
Storage temperature |
-20...70°C |
Humidity |
< 80%, not condensing |
Standards |
EN 60950:2006 + A11:2009 + A1:2010 + A12:2011 + AC:2011 + A2:2013EN 61236-1:2013-07 |
Overvoltage category |
2 |
Protection class |
1 |
Pollution degree |
2 |
Operational altitude |
< 2000 m |
Digital interfaces |
Featured |
1x USB-B for communication, 1x USB-A for functions, 1x LAN for communication |
Galvanic isolation from device |
Max. 400 VDC |
Terminals |
Rear side |
AC input, analog interface, USB-B, Ethernet |
Front side |
DC output, USB-A, remote sensing |
Dimensions |
Enclosure (WxHxD) |
276 x 103 x 355 mm |
Total (WxHxD) |
308 x max. 195 x min. 391 mm |
Weight |
≈ 7.5 kg |