Accurate measurements without the need to precisely position wire or cable in the center of the jaw clamp.SHOP NOW
The SDS3104XHD features an impressive array of measurement and math capabilities, options for a 50 MHz arbitrary waveform generator, as well as serial decoding. SHOP NOW
The Megger IMT100 is the most versatile multi-tester for PV systems, electric vehicles and industrial applications.SHOP NOW
Similar Items
Let us quote you our best price!
Allow us to earn your business and make this purchase easy.
0 reviews | Catalog: 24724 | Model: Sauermann LV 111 S
Portable thermo-anemometer with remote Diameter 14 mm (.55 inch ) vane probe, 6.5' (2m) cable (0.8 to 25 m/s and 0 to +50 Degrees C). 2-line display. Functions: air velocity, airflow in duct, airflow with cones. temperature, automatic average, Hold, Min, Max, backlight, change of unit and configurable automatic switch-off. Supplied with soft case and adjustment certificate.
List Price: $495.00
We accept purchase orders and all major credit cards. We also provide financing via Klarna at checkout for qualifying purchases.
Have a question or need to apply for payment terms?
Contact us.
Have a question? Ask an expert.
The Sauermann LV 111 S is a thermo-anemometer with an integrated Ø14 mm vane probe used to display velocity and temperature measurements.
Rotation of the vane probe leads to a circular magnet of 8 poles. A dual Hall effect sensor, placed next to the magnet captures the signals of magnetic field polarity transition. The sensor signal is converted to the electrical frequency and is proportional to the rotation velocity of the vane probe. Signal chronology allows us to determine the rotation direction.
Negative temperature coefficient probes are thermistors with a resistance that decreases with temperature according to the equation below:
RT = resistance sensor value at temperature T R(T0) = resistance sensor value at reference temperature T0 T and T0 in °C α and T0 sensor specific constants