GE Druck PACE CM (Control Module) Specifications |
Performance |
PACE CM0 Standard Precision |
0.02% Rdg + 0.02% FS (25 mbar: 0.20% rdg + 0.20% FS, 70 mbar: 0.10% rdg + 0.10% FS, 200 mbar: 0.04% rdg + 0.04% FS) includes linearity, hysteresis, repeatability and temperature effects for gauge pressures and assumes steady state temperature and regular zeroing. |
PACE CM0 Controller stability |
0.005% FS |
PACE CM1 High Precision |
0.01% Rdg + 0.01% FS (25 mbar: 0.10% rdg + 0.10% FS, 70 mbar: 0.05% rdg + 0.05% FS, 200 mbar: 0.02% rdg + 0.02% FS) includes linearity, hysteresis, repeatability and temperature effects for gauge pressures and assumes steady state temperature and regular zeroing. |
PACE CM1 Controller stability |
0.003% FS (25mbar range = 0.005% FS) |
PACE CM2 Premium Precision |
0.005% Rdg + 0.005% FS (25 mbar: 0.05% rdg + 0.05% FS, 70 mbar: 0.025% rdg + 0.025% FS, 200 mbar: 0.01% rdg + 0.01% FS) includes linearity, hysteresis, repeatability and temperature effects for gauge pressures and assumes steady state temperature and regular zeroing. |
PACE CM2 – Control stability |
0.001% FS (25mbar = 0.004% FS. 70 mbar = 0.003% FS) |
PACE CM2-A Aeronautical |
Altitude range -3000 to +55,000 ft:
Altitude precision @ Sea level ±2 ft @ 8500 ft ±3 ft @ 35,000 ft ±9 ft
Altitude RVSM accuracy: @ Sea level ±5 ft @ 29,000 ft ±25 ft @ 41,000 ft ±46 ft @ 35,000 ft ±33 ft
Airspeed range to 650 knots:
Airspeed precision @ 50 knots ±1.00 kts @ 250 knots ±0.21 kts @500 knots ±0.11 kts
Pressure Range –1/+1bar g, pressure precision 0.005% Rdg + 0.005% FS includes linearity, hysteresis, repeatability and temperature effects for gauge pressures and assumes steady state temperature and regular zeroing.
1300 mbar reference precision 0.005% FS, Includes non-linearity, hysteresis, repeatability and temperature effects between 15°C (59°F) and 45°C (113°F). |
PACE CM Long term stability measurement |
2 bar g to 210 bar g (30 psi to 3000 psi) 0.01% reading per annum, 1 bar g 0.02% reading per annum & 25 mbar g to 700 mbar g 0.03% reading per annum, assumes regular Zeroing. Barometric reference sensor 0.1 mbar a or 0.001450 psi a per annum (for CM0-B, CM1-B, CM2-B & CM2-A) |
PACE CM0-B Precision- Barometric Reference |
Precision for the optional barometric reference 0.10 mbar or 0.001450 psi. Includes non-linearity, hysteresis, repeatability and temperature effects between 15°C (59°F) and 45°C (113°F). |
PACE CM1-B Precision- Barometric Reference |
Precision for the optional barometric reference 0.05 mbar or 0.000725 psi. Includes non-linearity, hysteresis, repeatability and temperature effects between 15°C (59°F) and 45°C (113°F). |
PACE CM2-B Precision- Barometric Reference |
Precision for the optional barometric reference 0.025 mbar or 0.0003625 psi. Includes non-linearity, hysteresis, repeatability and temperature effects between 15°C (59°F) and 45°C (113°F). |
Weight |
5 Kg or 11 lbs |
Pressure Connection |
G 1/8 Female (1/8 NPT Female by adaptor for North America) |