GE Druck DPI 705 General Specifications |
70 or 200 mbar (1 or 3 psi) |
d |
|
350 mbar (5 psi) |
|
g, a, d |
1 bar (15 psi) |
d |
g, a, d |
2 bar (30 psi) |
g, a |
g, a, d |
3.5 bar (50 psi) |
|
g, a, d |
7 bar (100 psi) |
g |
g, a, d |
10 bar (150 psi) |
|
g, a, d |
20 bar (300 psi) |
g |
g, a, d |
35 bar (500 psi) |
|
g, a, d |
70 bar (1000 psi) |
|
g, a |
135, 200, 350 or 700 bar
(2000, 3000, 5000 or 10,000 psi) |
|
a |
a = absolute, d = differential (line pressure: 2 bar (30 psi) for internal ranges and
35 bar (500 psi) for remote sensors), g = gauge
Do not apply pressure greater than 1 bar (16 psi) absolute to the negative port
of differential pressure indicators. |
Accuracy |
±0.1% FS non-linearity, hysteresis and repeatability |
Temperature
Performance Averaged
32°F to 122°F
(0°C to 50°C) |
- ±0.01% rdg/°F (±0.02% reading/°C)
- For absolute ranges < = 15 psi (1 bar) add ±0.025% FS/°F (±0.05% FS/°C)
- For absolute ranges < 15 psi (1 bar) add ±0.01% FS/°F (±0.02% FS/°C)
- Calibrated range: 32°F to 122°F (0°C to 50°C)
- Operating range: 15°F to 122°F (-10°C to 50°C)
|
Humidity |
0 to 95% non condensing |
Overpressure |
Safe to 2 x FS. Display flashes at 110% FS |
Media Compatibility |
Stainless steel, except positive differential non-corrosive fluid and negative differential dry gas. |
Safety Conformance |
EN61010, EN50081-1, EN50082-2, CE marked |
Intrinsically Safe Version |
• IS Class I, Groups A, B, C & D, T4 Exia; Class I, Zone 0, AEx/Ex ia IIC T4 Ga (-10°C ≤ Ta ≤ +50°C) to CSA Std. 157-92, UL 913 (7th Ed.), CAN/CSA-C22.2 No. 60079-0:11, CAN/CSA-C22.2 No. 60079-11:11, ANSI/ UL 60079-0:09 and ANSI/UL 60079-11:09.
• Ex ia IIC T4 Ga (-10°C ≤ Ta ≤ +50°C) to ATEX IEC 60079-0:2011 and EN 60079-11:2012 |
Display |
13mm LCD with 99999 readout |
Weight |
<500 g |
Dimensions |
190 x 90 x 36 mm |
Battery |
3 x Alkaline cells |