Toll Free: (877) 571-7901
Local: (732) 728-2590
[email protected]
My Account
Sign in to see your member price, rewards, orders, tracking and more.
Why Join? Gain access to exclusive pricing, rewards and more!
Member Pricing & Rewards active
(877) 571-7901
Member Pricing & Rewards are active.
Accurate measurements without the need to precisely position wire or cable in the center of the jaw clamp.SHOP NOW
The SDS3104XHD features an impressive array of measurement and math capabilities, options for a 50 MHz arbitrary waveform generator, as well as serial decoding. SHOP NOW
The Megger IMT100 is the most versatile multi-tester for PV systems, electric vehicles and industrial applications.SHOP NOW
Similar Items
Let us quote you our best price!
Allow us to earn your business and make this purchase easy.
0 reviews | Catalog: 00-HTT-1R LomA | Model: Compliance West HTT-1R LomA
Function Tester, Ground (Pass 0.1 Fail 0.025 ohm) Leakage (Pass 4.232M, Fail 2.088M) Arcover 750V 400V
List Price: $490.00
Your Price: $477.27
Sign-in for EDU Pricing!
We accept purchase orders and all major credit cards. We also provide financing via Klarna at checkout for qualifying purchases.
Have a question or need to apply for payment terms?
Contact us.
Have a question? Ask an expert.
HTT-1R LomA, Function Checker, Ground (Pass 0.1 Fail 0.025 ohm) Leakage (Pass 4.232M, Fail 2.088M) Arcover 750V 400V. The HTT-1R provides a fast, safe and reliable way to verify proper operation of Dielectric Withstand (hipot) and Ground Continuity/Ground Bond Testers between calibration cycles, in accordance with requirements of TUV and other European agencies. It uses actual loads to simulate barely passing and barely failing results to exercise the tester(s), and is designed to be used as a verification test fixture in compliance with the requirements for shift by shift verification stated in ECS CIG-023 Inspection Report, Section 4. See the Manuals Section below for an example of the CIG-023 Inspection Report. It allows the operator to check the result of the three failure modes encountered in production hipot and ground circuit testing (open ground, high leakage current, and dielectric breakdown) separately or in any combination to verify proper test equipment operation. A passing test is also simulated. The test leads allows connection of the HTT-1R to any hipot tester using a standard NEMA 15A receptacle, or use the leads on the hipot and ground tester to connect directly to the HTT-1R. The manual contains anticipated results and suggests testing procedures. The maximum voltage the HTT-1R can test is 3000Vac/4200Vdc.
HTT-1R Front Panel Features
This procedure will describe the connection of the HTT-1R to a hipot tester or a combination hipot/ground bond tester. Because of the varying types of tester outputs, cables provided with the HTT-1R may not fit the hipot / ground bond tester. For custom cables, please contact Compliance West USA.
Connection to a hipot tester
This procedure will describe connection of the HTT-1R to a ground bond tester. Because of the varying types of tester outputs, only return banana to banana cable is provided with the HTT-1R that can be used on a Ground bond tester. For custom cables, please contact Compliance West USA. If the ground bond tester has an output current above 15A, make sure the test time is not higher than 3 seconds, to repeat another test wait for 10 seconds. Resistors inside may overheat, this can cause permanent damage and miss readings.