HT-10kVac: to 10,000Volts ac. Leakage current 1 - 10mA AC. HT-10kVdc: to 14,000Volts dc. Leakage current 1 - 5mA DC. HT-10kvac/dc: combines outputs and adds a current meter to the front panel. 3U enclosure. : Current capabilities of 25, 100 and 500mA available. Consult factory. Leakage Current: Adjustable via knob to any intermediate setting. Arc Detection: Separate high frequency detection circuit for breakdown spike detection. Test Time: 1-60 sec., defeatable for operation while button is pressed. Voltage Ramp Time: .5-5 seconds, defeatable. Leakage Fail Point: AC - 1 - 10mA DC - 1-5mA via knob. Voltage Ramp Defeat: applies voltage without ramp. Timer Duration Defeat: applies voltage only while TEST button is pressed. Shutdown on Failure Defeat: disables Arc Detection circuit. Voltage rampdown time: AC - .5 sec max; DC - 8 sec max. Duty Cycle: 100% Meter accuracy: ?3% 500-10kV |