Specifications |
Voltage Test Type |
AC |
Output Voltage |
0 - 10,000VAC |
Leakage Current Adjustment |
1 - 20 mA AC, increments of 0.1 mA |
Input Voltage |
120 V *different line voltages available |
Frequency |
50/60 Hz |
Fuse |
250 VAC 2 A Slow Blow |
Breakdown Detect |
Independent High Frequency Circuit |
Transformer Power |
< 500 VAC |
Leakage Current Detect |
Independent Low Frequency Circuit |
Test Time Range |
1 - 60 seconds |
Displays |
4 Digits LED Display (Voltage), 3 Digits LED Display (Current) |
Voltage Meter Accuracy |
± 2 % |
Current Meter Accuracy |
± 0.1 mA from 1 – 17mA ± 0.2 mA from 17 – 20mA |
Duty Cycle |
100 % |
Operation Temperature |
15 - 40 °C |
Relative Humidity Range |
0 - 90 % Non-Condensing |
General |
Dimensions |
17” Wide x 6” High x 17” Deep |
Weight |
32 lbs |
Product Package |
HT-10KVPac 20mA -HT-10KVPac 20mA Product Manual-18 AWG Test Lead (Black)-18 AWG High Voltage Test Lead (Red)-NIST Traceable Calibration Certificate to ANSI Z540 |
Options |
(xxx) Different Line Voltages |
Replace on the part number xxx-for 100, 110, 220, 230 or 240 depending the-line voltage. (120V for standard configuration). |
(XI) Safety Interlock |
Two position terminal block on rear panel-must be shorted for high voltage operation. |
(VF) Voltage on Failure |
If failure as Excess Leakage or Hipot-Fail happens, it will terminate the test and-the front voltage display-will hold the last voltage reading. |
(AR) Auto Reset on Pass |
Resets the tester after a few seconds only on passing test condition. |
(ZOS) Zero Voltage Start |
Voltage adjust knob must be to minimum position to allow start the test cycle.(RI) Relay Isolated Remote |
Interface |
It allows the interface with a PLC or other machine using relay closures. This option allows isolation between the PLC and the Hipot tester. |
Safety |
Visual Indicator of Voltage |
Provided by front panel meter, directly connected to high voltage output. |
Output |
Failure Indicator |
Audible, provided by buzzer.-isual, Provided by red LED's.-Test automatically terminated on failure. |
Leakage Test |
Provided, 1 mA AC factory set pass/fail point, user adjustable. |