EN45502-2-2 Sec. 20, External Defib Protection Test 1
A Compliance West surge Tester. That means it outputs a guaranteed waveform but won't break the bank. The tester outputs the waveform it is supposed to, time after time.
Built in the USA. Email and phone support are free for life. Reasonable repair and calibration charges. Timely and helpful customer service always.
The Defib Surge P 380 outputs the pulse shown in EN45502-2-2 Fig. 107; with a Tp of 1.5 – 2.5ms; and a Tw50 of 3 - 5.5ms when discharged into the resistor network shown in Fig. 110. For tester protection, Fig. 110’s R7 and R1 are supplied internally. In accordance with the standard, the Megapulse Defib Surge P 380 can recharge and trigger the next pulse within 20 seconds.
Options:
- Option RI: PLC Interface Board
- Option TMM: TestMINDER automation software
- Option XI: External interlock.
- Option Resistorn network; Built In resistor network of Figure 110.
- Option: 220, 230 or 240V Mains power