Comet T7410 Specifications |
Technical Parameters |
Value |
Output |
RS485 |
Measured Value |
Temperature + Relative Humidity + Atmospheric Pressure |
Construction Type |
Ambient Air |
Design |
Industrial |
Temperature Measuring Range |
-30 to 80 °C |
Relay Output |
No |
Two-State Input |
No |
Lcd Display |
Yes |
PoE |
No |
Relative humidity range |
0 to 100% |
Accuracy of relative humidity measurement |
±2.5% relative humidity from 5 to 95% at 23°C |
Accuracy of temperature output |
±0.4°C |
Available temperature units |
degrees Celsius, Fahrenheit |
Accuracy and range of dew point temperature output - for more details see graphs |
±1.5°C at ambient temperature T<25°C and RH>30%
range -60 to +80 °C |
Accuracy and range of absolute humidity output |
±1.5g/m3 at ambient temperature T < 25°C
range 0 to 400 g/m3 |
Accuracy and range of specific humidity output |
±2g/kg at ambient temperature T < 35°C
range 0 to 550 g/kg |
Accuracy and range of mixing ratio output |
±2g/kg at ambient temperature T < 35°C
range 0 to 995 g/kg |
Accuracy and range of specific enthalpy output |
±3kJ/kg at ambient temperature T < 25°C
range: 0 to 995 kJ/kg |
Maximum measuring pressure range |
600 to 1100hPa |
Accuracy of pressure measurement |
±1,3hPa 23°C from 800 to 1100hPa |
Available pressure units |
hPa, kPa, mbar, mmHg, inHg, inH2O, PSI, oz/in2 |
Temperature operating range |
-30 to +80°C |
Temperature operating range of LCD display |
readable to operating temperature +70°C, it is recommended to switch OFF the LCD over +70°C |
Temperature compensation of the humidity sensor |
all temperature range |
Communication protocol |
ModBus RTU and Advantech ADAM compatible protocol |
Communication speed |
110 to 115200 Bd |
Filtering ability of sensor cover |
0.025mm - filter with stainless steel mesh |
Protection of the case with electronics |
IP54 electronics, IP40 sensors |
Power |
9-30Vdc, consumption approximately 0.5W |
Dimensions |
88.5 x 170 x 39.5 mm (W x H x D), stem length 75 mm |
Weight |
approximately 150g |
Warranty |
3 years |