Comet T3319P Specifications |
Technical parameters |
Value |
Output |
RS232 |
Measured Value |
Temperature + Relative Humidity |
Construction Type |
With T+RH Probe on Cable |
Design |
Industrial |
Temperature Measuring Range |
-30 to 105 °C |
Relay Output |
No |
Two-State Input |
No |
Lcd Display |
Yes |
PoE |
No |
Relative humidity range |
0 to 100% |
Accuracy of relative humidity measurement |
±2.5% relative humidity from 5 to 95% at 23°C |
Accuracy of temperature output |
±0.4°C from -30 to +100°C ±0.4% from reading over +100°C |
Available temperature units |
degrees Celsius, Fahrenheit |
Accuracy and range of dew point temperature output - for more details see graphs |
±1.5°C at ambient temperature T<25°C and RH>30% range -60 to +80 °C |
Accuracy and range of absolute humidity output |
±1.5g/m3 at ambient temperature T < 25°C range 0 to 400 g/m3 |
Accuracy and range of specific humidity output |
±2g/kg at ambient temperature T < 35°C range 0 to 550 g/kg |
Accuracy and range of mixing ratio output |
±2g/kg at ambient temperature T < 35°C range 0 to 995 g/kg |
Accuracy and range of specific enthalpy output |
±3kJ/kg at ambient temperature T < 25°C range: 0 to 995 kJ/kg |
Temperature operating range of case with electronics |
-30 to +80°C |
Temperature operating range of LCD display |
readable to operating temperature +70°C, it is recommended to switch OFF the LCD over +70°C |
Range of humidity sensor temperature compensation |
all temperature range |
Communication protocol |
ModBus RTU and Advantech ADAM compatible protocol |
Communication speed |
110 to 115200 Bd |
Filtering ability of sensor cover |
0.025mm - filter with stainless steel mesh |
Protection of the case with electronics |
IP65 electronics, IP40 sensors |
Power |
9-30Vdc, current consumption approximately 6mA |
Length of the probe cable |
1m, 2m or 4m optionally |
Dimensions |
display unit 88.5 x 114 x 39.5 mm (W x H x D), probe length 88 mm, probe diameter 18mm |
Weight |
approximately 290g - 1m probe |
Warranty |
3 years |