| Gases Measured | R-11, R-12, R-22, R-23, R-113, R-114, R-123, R-124, R-134a, R-227, R-236fa, R-401A, R-402A, R-402B, R-404A, R-407A, R-407C, R-408A, R-409A, R-410A, R-500, R-502, R-503, R-507, R-508B, H-1211, H-1301, H-2402, N-1230 |
| Sensor | Infrared |
| Measurement Units | oz/yr, g/yr, mL/s, PaM3/s, ppm |
| Measurement Range & Accuracy | 0-50 oz/yr, ±10% of reading 50-1000 oz/yr, ±25% of reading |
| Warm-Up Time | 2 minutes |
| Probe | 8" (20.3 cm) flexible probe with 6' (1.8 m) or optional 12' (3.6 m) hose |
| Operating Modes | Search -- detects presence and location of gas leaks; Measure - calculates and displays leak rate |
| Response Time | Less than 1 second (with 6' hose) |
| Sensitivity | 0.01 oz/yr for most gases (minimum of 0.03 oz/yr for all gases) |
| Temperature Drift | ±0.8% of reading per °C |
| Calibration | Internal R-134a leak source, or External using customer supplied leak source |
| AC Input | 100?240 VAC, 50/60 Hz |
| Power Consumption | 15 Watts |
| Operating Temperature Range | 32 to 122°F (0 to 50°C) |
| Storage Temperature Range | ?4 to 122°F (-20 to 50°C) |
| Weight | 18 lbs (8.2 kg) |
| Dimensions | 4"H x 10.75"W x 15"D (10.2 x 27.3 x 38.1 cm)
|