Specifications |
Product Attributes |
Sensor |
Proprietary non-dispersive infrared (NDIR) technology |
Dimensions |
12.23" x 13.7" x 4.96" (21.0642 cm x 34.7980 cm x 12.5984 cm) |
Weight |
15 lbs (6.8 kg) |
User Interface |
Front panel with 3 indicator lights: Green - power on, normal; Yellow - fault; Yellow Flashing - system fault; Red Flashing - point has exceeded alarm set |
Communications |
Full 2-way communication with MZ-RD display module or building management system via RS-485 Serial Interface. RS-232C Comm. Port Standard00 |
Alarms |
Four SPDT alarm contacts are provided rated 2A at 250 VAC (inductive), 5A at 250 VAC (resistive). Three assigned to ppm level alarms, one assigned to system faults |
Conditioned Signal |
Optional dual 4-20 mA DC isolated outputs. Channel 1 = zone area, Channel 2 = ppm |
System Noise |
Less than 40dB at 10 ft (3m) |
Response Time |
5 to 315 seconds per zone - depending on air line length and number of zones |
Sampling Mode |
Automatic or manual (hold) |
Monitoring Distance |
1,200 ft max (500 ft for NH3) for combined length of sample and exhaust tubing (each zone) |
Power Safety Mode |
Fully automatic system reset. All programmed parameters retained |
Operating Temperature |
32° to 122° F (0 to 50° C) |
Ambient Humidity |
5% to 90% RH non-condensing |
Altitude Limit |
6,562 ft (2,000 m) |
Power |
100 to 240 VAC, 50/60 Hz, 20 W |
Approvals |
UL 61010-1, CAN/CA 22.2 No. 61010-1; EN61010-1, EN61326, EN14624; CE Mark |
Measurement |
Gas Library |
AGM-MZ: Ammonia (NH3), R717 |
Measuring Range |
AGM-MZ: Ammonia 25 to 10,000 ppm |
Accuracy |
AGM-MZ: ±10 ppm ±10% of reading from 0-1,000 ppm (most gases) |
Temperature Drift |
AGM-MZ: 1.5 ppm per degree C between purge cycles |