Bandwidth | 60 MHz |
Real time sample rate two channels interleaved | 400 MSa/s |
Equivalent sample rate | 40 GSa/s |
Rise time | < 5.83ns |
Channels | 2 |
Lower frequency limit | < 5Hz (at input BNC) < 1Hz when using a 10x passive probe |
Display | Color LCD, 5.7 inch (145mm) diagonal |
Record Length | 4000 points |
Vertical Resolution | 8 bit |
Vertical Sensitivity | 2mV - 5V/div |
DC gain accuracy | ±3.0% |
Maximum Input Voltage | 400 VPK, CAT II (between signal and reference BNC connector) |
Position Range | ± 8 divisions from center of screen |
Time Base Range | 2.5 ns/div - 50 s/div, |
Bandwidth Limit | 20 MHz |
Time Base Accuracy | 100 ppm |
Input Coupling | AC, DC, GND |
Input Impedance | 1MOhm || 19pF |
Vertical and Horizontal Zoom | Vertically or horizontally expand or compress a live or stopped waveform |
I/O interface | USB host port on front panel supports USB flash drives and USB printer interface supporting USB disk storage and USB printer USB device port for connection to PC. (Requires included Comsoft Software for use) |
Sample | Display sample data only |
Peak Detect |
Average | Waveform averaged, selectable from 2, 4, 16, 32, 64, 128, 256 |
Roll Mode | For time base settings 500 ms/div - 50s/div |
Trigger Types | Edge, Pulse Width, Video |
Trigger Modes | Auto, Normal, Single |
Trigger Coupling | AC, DC, LF reject, HF reject |
Trigger Source | CH1, CH2, AC line, Ext, Ext/5, Alternating |
Holdoff range | 100ns - 1.5s |
Trigger Level Range | CH1, CH2 | ± 8 divisions from center of screen | EXT | ± 1.6V | EXT/5 | ± 8V | |
Pulse Width Trigger | Pulse Width Range | 20ns - 10s | Trigger Condition | (>, <, =) positive Pulse (>, <, =) negative Pulse | |
Edge Trigger | Trigger Polarity | Rising and Falling Edge | |
Video | Supports formats PAL/SECAM, NTSC. Triggers on odd or even field, all lines or line number |
Types | Amplitude, Time |
Measurements | V, T, 1/T |
Time | Rise time, Fall Time, Frequency, Period, Positive Pulse width, Negative Pulse width, Delay, Phase, X at Max, X at Min |
Voltage | MAX, MIN, Peak-Peak, Average, Vrms, High, Low, Amplitude, Cycle RMS, Cycle Average, Overshoot, Preshoot |
Frequency | Hardware counter providing frequency readout of trigger source | Readout resolution | 5 digits | |
Math functions | FFT, add, subtract, multiply |
FFT | FFT Windows | Hanning, Hamming, Blackman, Rectangular, Flattop | Sample size | 2048 points | |
Single button automatic setup of both channels for vertical, horizontal and trigger systems |
Bandwidth | 60 MHz |
Phase Error | ±3.0 degrees |
Internal memory | 10 front panel setups 10 waveform files |
USB drive | Front panel setups, waveforms, bitmaps, csv files Store unlimited number of files (limited only by USB drive capacity) |
Display Mode | 1/4 VGA (320 × 240) 256 Color LCD with adjustable contrast and inverse video |
Display types | Point, Vector |
Persistence | Off, infinite |
Waveform interpolation | Sin(x)/x, Linear |
Format | YT and XY |
Input Voltage | 99 - 242 VAC, CAT II, Auto selection |
Frequency | 47Hz to 440Hz |
Power | 50VA Max |
Dimensions | 310mm (W) × 147mm (H) × 269mm (D) 12.2in (W) × 5.8in (H) × 10.6in (D) |
Weight | 3.6 kg (8 lbs) |
Temperature | Operating: 0°C to +40°C Nonoperating: -20°C to +55°C |
Humidity | Operating: 90%RH, 40°C Nonoperating: 90%RH, 65°C |
Altitude | Operating to 3000m |
Pollution Degree | Pollution degree 2 for indoor use only. |
EMC | This oscilloscope is in compliance with council EMC directive 2004/108/EC |
Safety | EN61010-1:2001 |
Supplied | User Manual, passive probe set (2 pieces), Power cord, USB interface cable, Comsoft Software Installation disk |
Optional | PR 37A 10:1 Probe, PR 32A Demodulator Probe, PR 55 High Voltage Probe |