Input Specifications |
Voltage |
115/230 V Auto Range, ± 15 % Variation |
Frequency |
50/60 Hz ± 5% |
Fuse |
115 VAC, 230 VAC – 10 A Slow Blow 250 VAC |
Dielectric withstand test mode |
Output Rating |
5 kV @ 50 mAAC
5 kV @ 100 mAAC (Models 825X)
6 kV @ 20 mADC |
Voltage Setting |
Resolution:
Accuracy: |
1 V
± (2% of setting + 5 volts |
HI and LO-Limit |
AC Total |
Range:
Resolution: |
0.000 – 9.999 mA
0.001 mA |
Range:
Resolution: |
10.00 – 50.00 mA (100.00 mA, models 825X)
0.01 mA |
Range:
Resolution: |
± (2% of setting + 2 counts) |
AC Real |
Range:
Resolution: |
0.000 – 9.999 mA
0.001 mA |
Range:
Resolution: |
10.00 – 50.00 mA (100.00 mA, models 825X)
0.01 mA |
Accuracy |
± (3% of setting + 50 μA) |
DC |
Range:
Resolution: |
0 – 999.9 μA
0.1 μA |
Range:
Resolution: |
1,000 – 20,000 μA
1 μA |
Accuracy |
± (2% of setting + 2 counts |
Arc Detection |
Range: |
1 – 9 (9 is most sensitive) |
Ground Continuity |
Current: DC 0.1 A ± 0.01 A, fi xed
Max. Ground Resistance: 1 Ω ± 0.1 Ω, fi xed |
Ground Fault Interrupt |
GFI Trip Current: 0.4 mA – 5.0 mA (AC or DC)
HV Shut Down Speed: < 1 ms |
DC Output Ripple |
≤ 4% Ripple rms at 5 kVDC at 20 mA Resistive Load |
Discharge Time |
≤ 50 ms No Load, < 100 ms for Capacitive Load |
Max Capacitive Load, DC Mode |
1 μF < 1 kV
0.75 μF < 2 kV
0.5 μF < 3 kV |
0.08 μF < 4 kV
0.04 μF < 6 kV |
AC Output Waveform |
Sine Wave, Crest Factor = 1.3 – 1.5 |
Output Frequency |
Range: |
60 or 50 Hz, User Selection (400/800 Hz optional) |
Output Regulation |
± (1% of output + 5 V) from no load to full load and over input voltage range |
Dwell Timer |
Range:
Range: |
AC 0.4 –999.9 sec (0=Continuous)
DC 0.3 –999.9 sec (0=Continuous |
Ramp Timer |
Ramp-up:
Ramp-Down: |
AC 0.1 – 999.9 sec, DC 0.4 – 999.9 sec
AC 0.0 – 999.9 sec, DC 0.0 , 1.0 – 999.9 sec (0=Continuous) |
Insulation Resistance test mode |
Voltage Setting |
Range: |
30 – 1000 VDC |
HI and LO-Limit |
Range:
Resolution: |
0.05 MΩ – 99.99 MΩ
0.01 MΩ |
Range:
Resolution: |
100.0 MΩ – 999.9 MΩ
0.1 MΩ |
Range:
Resolution: |
1,000 MΩ – 50,000 MΩ
1 MΩ (HI-Limit: 0=OFF) |
Ramp Timer |
Ramp-up:
Ramp-Down: |
0.1 – 999.9 sec
0.0, 1.0 – 999.9 sec (0=Continuous) |
Delay Timer |
Range: |
0.5 – 999.9 sec (0=Continuous) |