Run Test Mode (Models 82X6 & 82X7 only) |
DUT Power |
Voltage:
Current:
Range:
Resolution:
Accuracy |
0 – 277 VAC single phase unbalanced
16 AAC max continuous
0.0 – 277.0 VAC Full Scale
0.1 V
± (1.5% of reading +0.2 V), 30.0 – 277.0 VAC Short Circuit Protection: 23 AAC, Response Time < 3 sec |
Delay Time Setting |
Range: |
0.2 – 999.9 seconds |
Dwell Time Setting |
Range: |
0.1 – 999.9 seconds (0=Continuous) |
Trip Point Settings & Metering |
Voltage |
Volt-Hi
Volt-LO |
Range:
Resolution:
Accuracy: |
30.0 – 277.0 VAC
0.1 V
± (1.5% of setting + 0.2 V), 30.0–277 VAC |
Current |
Amp-HI
Amp-LO |
Range:
Resolution:
Accuracy: |
0.0 – 16.00 AAC
0.01 A
± (2.0% of setting + 2 counts) |
Watts |
Power-HI
Power-LO |
Range:
Resolution:
Accuracy: |
0 – 4,500 W
1 W
± (5.0% of setting + 3 counts) |
Power Factor |
PF-HI
PF-LO |
Range:
Resolution:
Accuracy: |
0.000 – 1.000
0.001
± (8% of setting + 2 counts) |
Leakage Current |
Leak-HI
Leak-LO |
Range:
Resolution:
Accuracy: |
0.00 – 10.00 mA (0=OFF)
0.01 mA
± (2% of setting + 2 counts) |
Timer Display |
Range:
Resolution:
Accuracy: |
0.0 – 999.9 seconds
0.1 second
± (0.1% of reading + 0.05 seconds) |
Leakage Current Test Mode(Models 82X6 & 82X7 only) |
Dut Power |
Voltage:
Current: |
0 – 277 VAC
16 AAC max continuous |
Voltage Display |
Range:
Resolution:
Accuracy: |
0.0 – 277.0 VAC Full Scale
0.1 V
± (1.5% of reading +0.2 V), 30.0 – 277.0 VAC |
Short Circuit
Protection: |
23 AAC, Response Time < 3 s |
Reverse Power Switch |
Reverse polarity switch setting select ON/OFF/AUTO
ON: Reverse power
OFF: Normal
AUTO: Automatic Reverse Polarity |
Neutral Switch |
ON/OFF selection for single fault condition |
Ground Switch |
ON/OFF selection for Class I single fault condition |
Probe Setting |
Surface to Surface (PH – PL)
Surface to Line (PH – L)
Ground to Line (G – L) |
Touch Current High Limit (rms) |
Range: Resolution: |
0.0 μA ~ 999.9 μA 1000 μA ~ 10.00 mA
0.1 μA / 1 μA / 0.01 mA |
Touch Current Display (rms) |
Range 1: |
0.0 μA ~ 32.0 μA, frequency DC, 15 Hz – 1 MHz |
Range 2: |
28.0 μA ~ 130.0 μA, frequency DC, 15 Hz – 1 MHz |
Range 3: |
120.0 μA ~ 550.0 μA, frequency DC, 15 Hz – 1 MHz |
Resolution for Ranges 1, 2, 3: |
0.1 μA |
Accuracy for Ranges 1, 2, 3: |
DC: 15 Hz < f <100 KHz: ± (2% of reading + 3 counts) 100 KHz < f < 1 MHZ: ± 5% of reading (10.0 μA – 999.9 μA) |
Range 4: |
400 μA ~ 2100 μA, frequency DC, 15 Hz – 1 MHz |
Range 5: |
800 μA ~ 8500 μA, frequency DC, 15 Hz – 1 MHz |
Resolution for Ranges 4 & 5: |
1 μA |
Accuracy for Ranges 4 & 5: |
DC: 15 Hz < f <100 KHz: ± (2% of reading + 3 counts) 100 KHz < f < 1 MHZ: ± 5% of reading (10 μA – 8500 μA) |
Range 6: |
8.00 mA ~ 10.00 mA, frequency DC 15 Hz – 100 kHz |
Resolution: |
0.01 mA |
Accuracy |
DC: 15 Hz < f < 100 KHz: ± 5% of reading (0.01 mA -10.00 mA) |
Touch Current Display (Peak) |
Range 1: |
0.0 μA ~ 32.0 μA, frequency DC – 1 MHz |
Range 2: |
28.0 μA ~ 130.0 μA, frequency DC – 1 MHz |
Range 3: |
120.0 μA ~ 550.0 μA, frequency DC – 1 MHz |
Resolution for Ranges 1, 2, 3: |
0.1 μA |
Accuracy for Ranges 1, 2, 3: |
DC: ± (2% of reading + 2 μA) 15 Hz < f < 1 MHZ : ± 10% of reading + 2 μA |
Range 4: |
400 μA ~ 2100 μA, frequency DC – 1 MHz |
Range 5: |
1800 A ~ 8500 μA, frequency DC – 1 MHz |
Resolution for Ranges 4 & 5: |
1 μA |
Accuracy for Ranges 4 & 5: |
DC: ± (2% of reading + 2 μA)
15 Hz < f < 1 MHz: ±(10% of reading + 2 μA) |
Range 6: |
8.0 mA ~10.00 mA, frequency DC – 100 KHz |
Resolution: |
0.01 mA |
Accuracy |
DC: ± (2% of reading + 3 counts) 15 Hz < f < 100 KHz: ± (10% of reading + 2 counts) |
MD Circuit Module |
MD1: UL544NP, UL484 , UL923, UL471, UL867, UL697
MD2: UL544P
MD3: IEC 60601-1
MD4: UL1563
MD5: IEC60990 Fig4 U2, IEC 60950-1, IEC60335-1,
IEC60598-1, IEC60065, IEC61010
MD6: IEC60990 Fig5 U3, IEC60598-1
MD7: IEC60950, IEC61010-1 FigA.2 (2K ohm) for Run function
MD8: IEC60990/60950 Fig4 U1 |
External MD |
Basic measuring element 1 kΩ |
Scope Output Interface |
BNC type connector on rear panel for Oscilloscope connection |