DC Withstand Test Mode (Models 7800/7804/7850 & 7854 Only) |
Output Voltage |
Range:
Resolution:
Accuracy: |
0 - 5,000 VAC
1 VAC
± (2% of setting + 5V) |
DC Output Ripple |
<4% (6 KV/10 mA at Resistive Load) |
HI and LO-Limit |
Range:
Resolution:
Accuracy: |
0.0000 – 0.9999 μA
0.0001 μA
± (2% of setting + 10 counts), Low Range is ON |
Range:
Resolution:
Accuracy: |
1.000 - 9.999 μA
0.001 μA
± (2% of setting + 10 counts), Low Range is ON |
Range:
Resolution:
Accuracy: |
10.00 - 99.99 μA
0.01 μA
± (2% of setting + 10 counts), Low Range is ON |
Range:
Resolution:
Accuracy: |
100.0 – 999.9 μA
0.1 μA
± (2% of setting + 2 counts) |
Range:
Resolution:
Accuracy: |
1,000 - 20,000 μA range (7804/54)
1,000 - 10,000μA range (7800/50)
1 μA
± (2% of setting + 2 counts) |
Ramp Up Timer |
Range: |
0.4 - 999.9 sec, Low Range is OFF
0.5 - 999.9 sec, Low Range is ON |
Ramp Down Timer |
Range: |
0.0, 1.0 - 999.9 sec (0=OFF) |
Dwell Timer |
Range: |
0, 0.4 - 999.9 sec (0=Continuous)
0, 1.0 - 999.9 sec, Low Range is ON |
Ramp-HI Selectable |
Range: |
0 - 20 mA selectable |
Charge-LO |
Range: |
0.0 - 350.0 μA DC or Auto Set |
Discharge Time |
< 50 ms for no load, < 100 ms for capacitive load |
Maximum
Capacitive Load
DC Mode |
1μF < 1kV
0.75 μF < 2 kV
0.5 μF < 3 kV |
0.0 μF < 4 kV
0.04 μF < 5 kV
0.015 μF < 6 kV |
Arc Detection |
Range:
Resolution:
Accuracy: |
1 - 9 ranges (9 is most sensitive) |