Input Specifications |
Voltage |
115/130 VAC ± 10%, Single Phase, User Selection |
Frequency |
50/60 Hz ± 5% |
Fuse |
6.3 A, 250 V Slow Blow |
Dielectric withstand test mode |
Output Rating |
07720 |
12 kV @ 10 mAAC |
HI-Limit and LO-Limit |
07720 |
Range 1:
Resolution:
Range 2:
Resolution: |
0.0 - 999.9 μA
0.1 μA
1,000 - 5,000 μA
1 μA/step |
077XX |
Accuracy: |
± (2% of setting + 2 counts |
DC Ramp Hi |
07720 |
6.75 mA peak maximum, 5 mADC, ON/OFF selectable |
DC Charge LO |
07710/07720 |
Range: |
0.0 - 350 μADC or auto set |
Arc Detection |
07710/07720 |
1 - 9 |
Voltage Display |
07715/07720 |
Range:
Accuracy: |
0.00 - 12.00 kV Full scale
± (2% of reading + 20 V) |
Current Display |
07720 |
Auto Range
Range 1:
Range 2: |
0.0 - 350.0 μA
300 - 5,000 μA |
DC Ouput Ripple |
07720 |
< 5% Ripple at 20 kV @ 4,999 μA, Resistive Load |
AC Output Waveform |
Sine Wave, Crest Factor = 1.3 - 1.5 |
Output Frequency |
Range: |
: 50/60 Hz, User Selection
± (1% of output + 5 V) from Regulation
No load to full load |
Output Regulation |
± (1% of output + 10 V) from no load to full load |
Discharge Timer |
07720 |
No load < 500 ms |
Dwell Timer |
Range:
AC Range:
DC Range: |
0, 0.3 - 999.9 sec (0=Continuous)
0, 0.3 - 999.9 sec or min (0=Continuous)
0, 0.4 - 999.9 sec or min (0=Continuous) |
Ramp Timer |
07710/07720 |
Range: |
0.4 - 999.9 sec |
Ground Continuity |
Max. Ground Resistance 1 Ω ± 0.1 Ω, fixed |
Ground Fault Interrupt |
HV Shut Down Speed < 1 ms
GFI Trip Current 1 mA max |
General Specifications |
Memory |
50 memories w/ 8 steps per memory |
Mechanical |
Tilt-up front feet |
Interface |
Standard: USB, RS-232 Optional: GPIB |
Dimensions (W x H x D) |
16.93" x 5.24" x 15.75" (430 x 133 x 400 mm) |