Input Specifications |
Voltage |
115/230 VAC ± 10%, User Selection |
Frequency |
50/60 Hz ± 5% |
Fuse |
2 A Slow Blow 250 VAC |
Line Conditions |
Reverse Power Switch |
Switch for power polarity reversal |
Neutral Switch |
Neutral switch on/off selection for single fault |
Ground Switch |
Ground switch on/off selection for class I single fault |
Probe Settings |
Surface to Surface |
(PH - PL) |
Surface to Line |
(PH - L) |
Ground to Line |
(G - L) |
Leakage Limit Settings |
Touch Current High/Low Limit (rms) |
Range:
Resolution: |
0.0 μA - 999.9 μA / 1,000 μA - 9,999 μA / 10.00 mA - 20.00 mA
0.1 μA / 1 μA / 0.01 mA |
Touch Current High/Low Limit (Peak) |
Range:
Resolution: |
0.0 μA - 999.9 μA / 1,000 uA - 9,999 μA / 10.00 mA - 30.00 mA
0.1 μA / 1 μA / 0.01 mA |
Display |
Touch Current Display (rms) |
Range:
Resolution:
Accuracy: |
0.0 μA - 550 μA, frequency DC, 15 Hz - 1 MHz
0.1 μA
DC: 15 Hz ≤ f ≤ 100 kHz: ± (2% of reading + 3 counts) 100 kHz ≤ f ≤ 1 MHz: ± 5% of reading (10.0 μA - 999.9 μA) |
Range:
Resolution:
Accuracy: |
400 μA - 8,500 μA, frequency DC, 15 Hz - 1 MHz
1 μA
DC: 15 Hz ≤ f ≤ 100 kHz: ± (2% of reading + 3 counts) 100 kHz ≤ f ≤ 1 MHz: ± 5% of reading, (10.0 μA - 8,500 μA |
Range:
Resolution:
Accuracy: |
8.00 mA - 20.00 mA, frequency DC, 15 Hz - 100 KHz
0.01 mA
DC: 15 Hz ≤ f ≤ 100 MHz: ± 5% of reading (0.01 mA - 20.00 mA) |
Touch Current Display (peak) |
Range:
Resolution:
Accuracy: |
0.0 μA - 550 μA, frequency DC - 1 MHz
0.1 μA
± (2% of reading + 2 μA)15 Hz ≤ f ≤ 1 MHz, ± 10% of reading + 2 μA |
Range:
Resolution:
Accuracy: |
400 μA - 8,500 μA, frequency DC - 1 MHz
1 μA
± (2% of reading + 2 μA) 15 Hz ≤ f ≤ 1 MHz, ± 10% of reading + 2 μA |
Range:
Resolution:
Accuracy: |
8.00 mA - 30.00 mA, frequency DC - 100 kHz
0.01 mA
± (2% of reading + 3 counts) 15 Hz ≤ f ≤ 100 kHz, ± 10% of reading + 2 counts |
Measuring device module |
MD1 |
UL544NP, UL484 , UL923, UL471, UL867, UL697 |
MD2 |
UL544P |
MD3 |
IEC 60601-1 |
MD4 |
UL1563 |
MD5 |
IEC60990 Fig4 U2, IEC60950-1, IEC60335-1, IEC60598-1,IEC60065, IEC61010 |
MD6 |
IEC60990 Fig5 U3, IEC60598-1 |
MD7 |
IEC60950, IEC61010-1 FigA.2 (2 kohm) for Run function |
External MD |
Basic measuring element 1 kohm |
MD Voltage Limit |
70 VDC |
DUT power |
AC Voltage |
0.0 - 277.0 V |
AC Current |
40 A max continuous |
AC Voltage High/Low Limit |
Range:
Resolution: |
0.0 - 277.0 V
0.1 V/step |
AC Voltage Display |
Range:
Resolution: |
0.0 - 277.0 V
0.1 V/step
± (1.5% of reading + 2 counts), 30.0 – 277.0 V |
Delay Time Setting |
Range:
Resolution: |
0.5 - 999.9 sec
0.1 sec |
Dwell Time Setting |
Range:
Resolution: |
0, 0.5 - 999.9 sec (0=Continuous)
0.1 sec
± (0.1% of reading + 0.05 seconds) |
Failure Protection |
On Start-Up - Neutral Voltage Check (Neutral - V)
Over current and ground current check (Line - OC) |
General Specifications |
Memory |
50 Memories, 30 steps per each memory File locations can link 900 steps max |
Mechanical |
Bench or rackmount with tilt-up feet |
Interface |
Standard: USB, RS-232 Optional: Ethernet, GPIB |
Dimensions (W x H x D) |
16.93" x 5.24" x 11.81" (430 x 133 x 300 mm) |
Weight |
26.45 lbs (12 kg) |