PERFORMANCE SPECIFICATIONS |
Accuracy |
Total Error Band includes combined effects of temperature, non-linearity (Terminal Point Method), hysteresis, non repeatabilty, zero offset and span setting errors.
±1% of Span: From –20 to 85ºC (–4 to 185ºF)
±1.5% of Span: From –40 to –20ºC (–40 to –4ºF)
±1.5% of Span: From 85 to 125ºC (185 to 257ºF)
Note: Static accuracy ±0.25% of span BFSL (Best Fit Straight Line Method); includes non-linearity, hysteresis and non-repeatable effects at reference temperature 72°F (21°C) |
Stability |
Less than ±0.25% span/year |
Durability |
Tested to 50 million cycles |
ENVIRONMENTAL SPECIFICATIONS |
Temperature |
Compensated: –40 to 125°C (–40 to 257°F)
Operating: –40 to 125°C (–40 to 257°F)
Storage: –40 to 125°C (–40 to 257°F) |
Humidity |
0 to 100% R.H., no effect |
FUNCTIONAL SPECIFICATIONS |
Select from over 25 pressure ranges starting at 30 psi and running through 20,000 psi gauge. Compound (vacuum & pressure) ranges are also available, see “To Order” on back. |
Overpressure (F.S.) |
Proof |
Burst |
750 psi & below
1500 psi
3000 psi
5000 psi
7500 psi
10,000 psi
20,000 psi |
200% F.S.
200% F.S.
200% F.S.
150% F.S.
120% F.S.
120% F.S.
120% F.S. |
1000% F.S.
500% F.S.
500% F.S.
500% F.S.
500% F.S.
240% F.S.
240% F.S. |
Vibration |
Random vibration (20 g) over temperature range (–40° to 125°C). Exceeds typical MIL. STD. requirements |
Shock |
100gs, 6 ms |
Drop Test |
Withstands 1 meter on concrete 3 axis |
Response Time |
Less than 1 msec |
Warm-up Time |
Less than 500 msec typical |
Position Effect |
Less than ±0.01% span, typical |
ELECTRICAL SPECIFICATIONS |
Output Signals Available |
Voltage Output |
Excitation |
Supply Current |
0-5 Vdc, 3 wire
0-10 Vdc, 3 wire
1-5 Vdc, 3 wire
1-6 Vdc, 3 wire
0.5-4.5Vdc, 3 wire |
9-36 Vdc
14-36 Vdc
9-36 Vdc
9-36 Vdc
9-36Vdc |
5mA
5mA
4mA
4mA
4mA |
Ratiometric Output |
|
|
0.5-4.5 Vdc, 3 wire |
5 Vdc ±0.5 Vdc |
3.5mA |
Current Output |
|
|
4-20mA, 2 wire |
9-36 Vdc |
|
Reverse Polarity & Miswired Protected |
Yes |
Insulation Breakdown Voltage |
100 Vac |
Insulation Resistance |
Greater than 100 megohms at 100 Vdc |
CE Marked |
Per DoC
EMC Directive 2004/108/EC
IEC/EN 61326-1:Edition 1.0 Industrial
IEC/EN 61326-2-3:Edition 1.0 Annex BB Industrial
PED Directive |
PHYSICAL SPECIFICATIONS |
Wetted Materials |
304SS pressure connection and 17-4PH SS sensor diaphragm |
Housing |
20% Glass Reinforced Nylon, Fire retardant to UL94 V1 |
Available Process Connections (Male) |
See “How To Order” section below. For other connections consult factory |
Ingress Rating |
• IP67, NEMA 4X:
- Metri-Pack I50 series*
- Shielded cable
- Flying leads
- Deutsch DT Series DT04-3P
- Deutsch DTM Series DTM04-3P
- AMP Superseal
- M12 - Mates to Hirschmann 933 172-100 or similar
• IP65, NEMA 4X:
- Hirschmann G series**
- EN 175301-803, Form C (DIN 43650, Form C) |
|
|
|