| PERFORMANCE SPECIFICATIONS |
| Ref. Temperature |
70°F ±2°F (21°C ±1°C) |
| Accuracy Class (Span) |
±0.8% or ±0.4% Includes non-linearity, hysteresis, non-repeatability, zero offset and span setting errors. |
| Stability – Max. Change (Span/year) |
≤0.25%, ≤0.25% |
Standard Ranges (Inches W.C.)
Unidirectional Ranges |
Differential or Gauge
0/0.10
0/0.25
0/0.50
0/0.75
0/1.00
0/2.00
0/2.50
0/3.00
0/5.00
0/10.00
0/15.00
0/25.00 |
Standard Ranges (Inches W.C.)
Bidirectional Ranges |
Compound
±0.10
±0.25
±0.50
±1.00
±2.00
±5.00
±10.00
±15.00 |
| Response Time |
250 msec |
| ENVIRONMENTAL SPECIFICATIONS |
| Temperature Limits |
Storage: –40 to 180°F
Operating: +0 to 160°F
Compensated Range: +35 to 130°F
(10-95% R.H. non-condensing) |
| Thermal Coefficients |
Zero & Span: ±0.03% Span/°F |
| CE Marked |
Per DoC
EMC Directive 2004/108/EC
IEC/EN 61326-1:Edition 1.0 Industrial
IEC/EN 61326-2-3:Edition 1.0 Annex BB Industrial |
| FUNCTIONAL SPECIFICATIONS |
| Overpressure Limits |
Proof Pressure: 15 psid
Burst Pressure: 25 psid
Max. static line pressure: 25 psi |
| Mounting Position Effect |
±1%/g (lowest range)
Note: Calibration in vertical position is standard. |
| ELECTRICAL SPECIFICATIONS |
| Voltage Output |
4-20mA (2 wire)
0-5 Vdc (3 wire)
0-10 Vdc (3 wire) |
| Supply Voltage |
12-36 Vdc
11.5-36 Vdc
14-36 Vdc or 24 Vac |
| Supply Current |
21.5mA
4.5mA
6mA |
Output signal is independent of power supply changes
Reverse Wiring Protected
Zero and Span Adjustment |
Externally accessible
Zero: ±5% F.S.
Span: ±5% F.S. |
| PHYSICAL SPECIFICATIONS |
| Pressure Connections |
1⁄4˝ brass barbed fittings
1⁄8 NPT Female brass |
| Electrical Connection |
Euro style pluggable terminal block accepts 12-26 gauge wire |
| LED visual indicator standard |
| Weight |
Approx. 2.5 oz |
| Media |
Clean, dry and non-corrosive gas |
| Mounting |
Threaded fastener and 35mm DIN rail mount standard |
| Option |
1⁄2˝ plenum/conduit mounting bracket and cover kit (order part #101A213-01) |
| XRH |
(9 point NIST Calibration Certification) |