Electrical Specifications 23 °C ± 5 °C, < 75% R.H. |
Function |
Range |
Accuracy |
DC VOLTS |
|
400 mV, 4 V, 40 V, 400 V, 600 V |
±(1.0% rdg + 2 dgts) |
Resolution: |
0.1 mV |
|
Input impedance: |
400 mV: >100 MΩ; 4 V:10 MΩ; 40 V ~ 600 V: 9.1 MΩ |
|
Overload protection: |
600VDC or AC rms |
|
AC VOLTS |
|
400 mV, 4 V, 40 V, 400 V, 600 V |
|
|
400 mV (50 ~ 100 Hz) |
±(2.0% rdg + 5 dgts) |
|
4, 40, 400, 600 V (50 ~ 500 Hz) |
±(2.0% rdg + 5 dgts) |
Resolution: |
0.1 mV |
|
Input impedance: |
400 mV: >100 MΩ; 4 V: 10 MΩ; 40 V ~ 600 V: 9.1 MΩ |
|
Overload protection: |
600 VDC or AC rms |
|
DC CURRENT |
|
400 μA, 4000 μA, 40 mA, 400 mA, 10 A |
|
|
400 μA to 400 mA ranges |
±(2.0% rdg + 2 dgts) |
|
10 A range |
±(3.0% rdg + 3 dgts) |
10A input: 10 A for 60 seconds maximum followed by a 10 minutes cooling period |
Resolution: |
0.1 μA |
|
Voltage burden: |
0.2 V on 400uA, 40mA ranges
2V on 4000 uA, 400 mA ranges |
|
Input protection: |
0.5 A / 500V fast blow ceramic fuse (FP520)
10 A / 500 V fast blow ceramic fuse (FP540) |
|
AC CURRENT (50-500 Hz) |
|
400 uA, 4000 uA, 40 mA, 400 mA, 10 A |
|
|
400 uA to 400 mA ranges |
±(2.5% rdg + 5 dgts) |
|
10 A range |
±(3.5% rdg + 5 dgts) |
10A input: 10 A for 60 seconds maximum followed by a 10 minutes cooling period |
Voltage burden: |
see DC Current |
|
Input protection: |
see DC Current |
|
RESISTANCE |
|
400 Ω, 4 kΩ, 40 kΩ, 400 kΩ, 4 MΩ, 40 MΩ |
|
Resolution: |
0.1 Ω on 400 Ω range |
|
|
400 Ω to 400 kΩ ranges |
±(1.5% rdg + 4 dgts) |
|
4 MΩ range |
±(2.5% rdg + 4 dgts) |
|
40 MΩ range |
±(5.0% rdg + 5 dgts) |
Open circuit volts: |
-0.45Vdc (-1.2Vdc on 400 Ω range) |
|
Overload protection: |
600 VDC or AC RMS |
|
FREQUENCY (Hz) (Auto ranging) |
|
4 kHz, 40 kHz, 400 kHz |
|
Resolution: |
1Hz |
±(0.1% rdg + 3 dgts) |
Sensitivity: |
10Hz ~ 400kHz: > 3.5Vrms |
|
Minimum pulse width: |
> 2.5us |
|
Duty cycle limits: |
>30% and < 70% |
|
Overload protection: |
600 VDC or AC RMS |
|
TEMPERATURE |
Ranges: |
-20°C ~ 1000°C, -4°F ~ 1832°F |
|
Resolution: |
0.1°C, 0.1°F |
|
|
-20°C ~ 10°C |
±(2.0% rdg + 4°C) |
|
10°C ~ 200°C |
±(1.0% rdg + 3°C) |
|
-20°C ~ 10°C |
±(2.0% rdg + 4°C) |
|
200°C ~ 1000°C |
±(3.0% rdg + 2°C) |
|
-4°F ~ 50°F |
±(2.0% rdg + 8°F) |
|
50°F ~ 400°F |
±(1.0% rdg + 6°F) |
|
400°F ~ 1832°F |
±(3.0% rdg + 4°F) |
Sensor type: |
K-type thermocouple |
|
Overload protection: |
600 VDC or AC RMS |
|
RPM |
Range: RPM |
600 ~ 4000 RPM |
|
X10 RPM |
4000 ~ 12000 RPM |
|
Resolution: |
1 RPM |
±(2% rdg + 4 dgts) |
Effect Reading: |
> 600 RPM |
|
Overload protection: |
600 VDC or AC RMS |
|
% DUTY CYCLE |
|
1.0% ~ 90.0% |
|
Resolution: |
0.1% |
|
Pulse width: |
> 100us, < 100ms |
±(2% rdg + 5 dgts) |
Overload protection: |
600 VDC or AC RMS |
|
DWELL ANGLE |
No. of Cylinders: 4, 5, 6, 8 |
|
|
|
4 CYL |
0 ~ 90.0° |
|
5 CYL |
0 ~ 72.0° |
|
6 CYL |
0 ~ 60.0° |
|
8 CYL |
0 ~ 45.0° |
Resolution: |
0.1° |
±(2% rdg + 5 dgts) |
Overload protection: |
600 VDC or AC RMS |
|
CONTINUITY |
Audible indication: |
Less than 25Ω |
|
Response time: |
500ms |
|
Overload protection: |
600VDC or AC RMS |
|
DIODE TEST |
Test current: |
1.0mA (approximate) |
±(3.0% rdg + 3 dgts) |
Resolution: |
10mV |
|
Audible indication: |
<0.25V |
|
Open circuit volts: |
3.0VDC typical |
|
Overload protection: |
600VDC or AC RMS |
|