Electrical Specifications |
Function |
Range |
Accuracy |
Voltage |
DC Volts |
Ranges |
200 mV, 2 V, 20 V, 200 V, 1000 V |
± (0.5 % rdg + 1 dgt) |
AC Volts (45 Hz to 500 Hz) |
Ranges |
200 mV (45 Hz to 100 Hz) |
± (1.5 % rdg + 5 dgts) |
|
2 V, 20 V, 200 V, 750 V |
± (1.5 % rdg + 5 dgts) |
Resolution |
100 μV |
|
Input Impedance |
400 mV: >100 MW; 4 V: 10 MW; 40 V to 1000 V: 9.1 MΩ |
|
Overload Protection |
1000 V DC or 750 V AC RMS |
|
Current |
DC Current |
Ranges |
200 μA, 2000 μA, 20 mA, 200 mA |
± (1.0 % rdg + 2 dgt) |
|
2 A |
± (2.0 % rdg + 3 dgts) |
AC Current (45 Hz to 500 Hz) |
Ranges |
200 μA, 2000 μA, 20 mA, 200 mA |
± (1.5 % rdg + 5 dgts) |
|
2 A |
± (2.5 % rdg + 5 dgts) |
Resolution |
0.1 μA |
|
Input Protection |
2 A/1000 V fast blow ceramic fuse 6.3 mm x 32 mm on A input |
|
Burden Voltage |
500 mV max. (2 V max. on 2 mA, 200 mA ranges) |
|
Resistance |
200 W, 2 kW, 20 kW, 200 kW, 2 MΩ |
± (1.0 % rdg + 4 dgts) |
|
2000 MΩ |
± {5.0 % (rdg -10 dgts) +20 dgt} |
Resolution |
100 mΩ |
|
Open Circuit Voltage |
-0.45 V DC typical, (-1.2 V DC on 200 Ω range) |
|
Overload Protection |
1000 V DC or 750 V AC RMS |
|
Continuity |
Audible Indication |
Less than 25 Ω |
|
Response Time |
100 ms |
|
Overload Protection |
1000 V DC or 750 V AC RMS |
|
Diode Test |
Test Current |
1.2 mA (approximate) |
±(1.5 % rdg + 3 dgts) |
Resolution |
1 mV |
|
Open Circuit Voltage |
3.0 dc typical |
|
Overload Protection |
1000 V DC or 750 V AC RMS |
|
Logic Test |
TTL |
|
Thresholds Logic 1 (Hi ) |
2.8 V ± 0.8 V |
|
Thresholds Logic 0 (Lo) |
0.8 V ± 0.5 V |
|
Test Voltage |
TTL 5 VDC |
|
Frequency Range |
20 MHz |
|
Pulse Width |
25 ns min |
|
Duty Cycle Limits |
> 30 % and < 70 % |
|
Indication |
40 ms beep at Logic 0 ( Lo) |
|
Overload Protection |
1000 VDC or 750 VAC RMS |
|
Non-Contact Voltage Indicator |
Sense voltage 70 V to 600 V ac (50 Hz to 60 Hz) beeper chirps and big bright red LED comes on, works on all ranges |