General Specifications |
Display |
4 digit LCD, 9999 counts, with annunciators, menu features and 41 segment bargraph. |
Polarity Indication |
Automatic |
Input overload |
indication OL, -OL. |
Low Battery Indication |
B; less than 50 hours battery life remain, accuracy is no longer guaranteed |
Display Update Rate |
2/sec, nominal; 20/sec for bargraph. |
Oper. Temp. |
0 °C to +50 °C @ 0 to 75 % R.H. |
Storage Temp. |
-20 °C to 60 °C @ 0 to 80 % RH, battery removed |
Altitude |
2000 meters - indoor use |
Temperature coefficient |
0.1 x (spec. accuracy)/°C (0 °C to 18 °C and 28 °C to 50 °C) |
Fuse |
F 2 A/1500 V fuse (8 mm x 65 mm), I.R. 30 kA – Amprobe p/n FP700 |
Power |
Standard 9-volt battery, NEDA 1604, JIS 006P, IEC 6F22 |
Auto Power-Down |
Meter powers down after approximately 30 minutes of inactivity. Not in Min/Max function. |
Battery Life (typical) |
150 hours, alkaline. Backlight usage consumes extra power and will decrease battery life significantly. Backlight auto-off after approx. 60 seconds. |
Dimensions, without holster
(H x W x D) |
200 x 102 x 59 mm (7.9 in x 4.0 in x 2.3 in) |
Weight (incl. battery) |
642 g (22 oz) |
Accessories |
Heavy Duty CAT IV 1000 V Test leads with threaded alligator clips (TL1500), battery (installed), hex wrench (inside holster), Holster with Magne-Grip strap, Type K thermocouple probe (TP255), Temperature Adaptor (TA--1A) and User Manual. |
Case material |
Reinforced, high-impact, fire retardant thermoplastic |
Safety |
Meets EN 61010-1 CATIV1000V(1500VDCmax)ClassII,EN60529:IP67CAT IV 1000 V (1500 V DC max) Class II, EN 60529:IP67 |
EMC |
Meets EN 61326-1 |
Electrical Specifications |
DC Volts Ranges |
1000 mV, 10 V, 100 V, 1500 V |
DC Volts Resolution |
0.1 mV in 1000 mV range |
DC Volts Accuracy |
±(0.1 % rdg + 5 dgt) |
DC Volts Input Impedance |
10 MΩ |
DC Volts CMRR |
>120 dB up to 1500 V dc |
DC Volts NMRR |
>60 dB at 50 or 60 Hz |
DC Volts OL Protection |
1500 V dc or 1000 V ac rms |
DC Volts Transient protection |
12 kV impulse (1.2 ìS/50 ìS) based on EN 61010-1:2001 impulse requirement for at CAT IV 1000 V/1500V dc product. This product should not be used in installations |
AC Volts True rms Ranges |
1000 mV, 10 V, 100 V, 1000 V |
AC Volts True rms Resolution |
0.1 mV in 1000 mV range |
AC Volts True rms Accuracy |
1000 mV (45 Hz to 400 Hz) ± (1.2 % rdg + 10 dgt) |
AC Volts True rms Accuracy |
10 V, 100 V (45 Hz to 500 Hz) ± (1.2 % rdg + 10 dgt) |
AC Volts True rms Accuracy |
10 V, 100 V (500 Hz to 2 kHz) ± (2.0 % rdg + 10 dgt) |
AC Volts True rms Accuracy |
1000 V (45 Hz to 1 kHz) ± (2.0 % rdg + 10 dgt) |
AC Volts True rms Input Impedance |
10 MΩ |
AC Volts True rms Conversion type |
True rms, ac coupled 5 % to 100 % of range |
AC Volts True rms Crest factor |
≤ 3 |
AC Volts True rms OL Protection |
1500 V dc or 1000 V ac rms |
Transient protection |
12 kV impulse (1.2 µS/50 µS) based on EN 61010-1:2001 impulse requirement for at CAT IV 1000 V product. This product should not be used in installations where transients exceed 12 kV |
DC Current Ranges |
100 µA, 1000 µA, 10mA, 100mA, 400mA, 2A (Auto/Manual ranging) |
DC Current Resolution |
0.01µA in 100 ±A range |
DC Current Accuracy |
100 µA range ± (0.5% rdg + 10 dgts) |
DC Current Accuracy |
1000 µA to 400 mA ranges ± (0.5% rdg + 5 dgts) |
DC Current Accuracy |
2 A range ± (1.5% rdg + 10 dgts) |
DC Current Input protection |
2 A/1500 V fast blow ceramic fuse 8X65 mm on A input , FP700 |