U1610A |
U1620A |
Specification |
Vertical system |
Bandwidth (-3 dB) 1 |
100 MHz |
200 MHz |
DC vertical gain accuracy 1 |
± 4% of full scale
Full scale is equivalent to 8 div |
Dual cursor accuracy 1 |
± {DC vertical gain accuracy + 0.4% full scale (~1 least significant bit (LSB)}
± {4% full scale ± 0.4% full scale (~1 LSB)} |
Characteristic |
Acquisition |
Maximum Sampling Rate
Single Chanel Operation
Dual Channel Operation |
1 GSa/s interleave
500 MS/s each channel |
2 GSa/s interleave
1 GS/s each channel |
Maximum Recording Length
Single Chanel Operation
Dual Channel Operation |
120 Kpts interleave
60 Kpts each channel |
2 Mpts interleave
1 Mpts each channel |
Vertical resolution |
8 bits |
Peak detection |
> 10 ns |
> 5 ns |
Average |
Selectable from 2 to 8192 in powers-of-2 increments |
Filter |
10 kHz and 20 MHz bandwidth limiters |
Interpolation |
(Sin x)/x |
Vertical system |
Analog channels |
Channel 1 and Channel 2 simultaneous acquisition |
Calculated rise time |
3.50 ns typical |
1.75 ns typical |
Vertical scale |
2 mV/div to 50 V/div |
Maximum input |
CAT III 600 Vrms (with 10:1 probe)
CAT III 300 Vrms (direct / 1:1 probe) |
Offset (position) range |
± 4 div |
Dynamic range |
± 8 div |
Input impedance |
1 MΩ ± 1% ˜ 22 pF ± 3 Pf |
Coupling |
DC, AC |
Bandwidth limit |
10 kHz and 20 MHz (selectable) |
Channel-to-channel isolation (with channels at the same V/div) |
CAT III 600 Vrms |
Probes |
U1560-60002 1:1 passive probe
U1561-60002 10:1 passive probe
U1562-60002 100:1 passive probe |
Probe attenuation factors |
1x, 10x, 100x |
Probe compensation output |
5 V pp , 1 kHz |
Noise peak-to-peak (typical) |
3% of full scale or 5 mV pp , whichever greater |
DC vertical offset (position) accuracy |
± 0.1 div ± 2 mV ±1.6% offset value |
Single cursor accuracy |
± {DC vertical gain accuracy + DC vertical offset accuracy + 0.2% full scale (~½ least significant bit (LSB)}
± {4% full scale ± 0.1 div ± 2 mV ± 1.6% offset value + 0.2% full scale (~½ LSB)} |
Horizontal system |
Range |
5 ns/div to 50 s/div |
2 ns/div to 50 s/div |
Resolution |
100 ps for 5 ns/div |
40 ps for 2 ns/div |
Timebase accuracy |
25 ppm |
Reference position |
Left, center, right |
Delay range (pre-trigger) |
1 screen width or 120 µs (whichever less) |
1 screen width or 1 ms (whichever less) |
Delay range (post-trigger) |
50 ms to 500 s |
20 ms to 500 s |
Delay resolution |
100 ps for 5 ns/div |
40 ps for 2 ns/div |
Delay time measurement accuracy |
Same channel: ± 0.0025% reading ± 0.17% screen width ± 60 ps
Channel-to-channel: ± 0.0025% reading ± 0.17% screen width ± 120 ps |
Modes |
Main, zoom, XY, roll |
Horizontal pan and zoom |
Dual window zoom |
Trigger system |
Sources |
Channel 1, Channel 2, External |
Modes |
Normal, Single, Auto |
Types |
Edge, Glitch, TV, Nth Edge, CAN, LIN |
Autoscale |
Finds or displays active channels, sets the edge trigger type on the highest numbered
channel, and sets the vertical sensitivity on the scope channel timebase to display
~2 periods
Requires > 10 mV pp minimum voltage, 0.5% duty cycle, and > 50 Hz minimum frequency |
Holdoff time |
60 ns to 10 s |
Range |
± 6 div from center of screen |
Sensitivity |
= 10 mV/div: 0.5 div
< 10 mV/div: greater of 1 div or 5 mV |
Trigger level accuracy |
± 0.6 div |
Coupling modes |
AC (~10 Hz), DC, LF-Reject (~35 kHz), HF-Reject (~35 kHz) |
External trigger |
|
1 MΩ ˜ 10 pF |
|
CAT III 300 Vrms |
|
DC coupling: trigger level ± 5 V |
|
100 kHz |
Measurement |
Automatic measurements |
Delay, duty cycle (+/–), fall/rise time, frequency, period, phase shift, T-max, T-min, width
(+/–), amplitude, average, base, crest, cycle mean, maximum, minimum, overshoot,
peak-to-peak, preshoot, standard deviation, top, Vrms (AC/DC), active/apparent/reactive
power, power factor |
Waveform math functions |
CH1 + CH2, CH1 – CH2, CH2 – CH1, CH1 × CH2, CH1/CH2, CH2/CH1, d/dt (CH1), d/dt
(CH2), ∫(CH1)dt, ∫(CH2)dt, FFT |
Cursors |
Delta V: Voltage difference between cursors
Delta T: Time difference between cursors |
FFT points |
1024 |
FFT windows |
Rectangular, Hamming, Hanning, Blackman-Harris, Flattop |
Display system |
Display |
5.7" TFT LCD VGA Color (outdoor readable) |
Resolution |
VGA (screen area): 640 vertical by 480 horizontal |
Control |
Vectors on/off, sin x/x interpolation on/off, infinite persistence on/off, backlight
intensity, color scheme, clear display |
Real-time clock |
Date and time (adjustable) |
Language |
10 languages (selectable) |
Built-in help system |
Functional quick help displayed by pressing the [Help] button |
Storage system |
Save/recall (non-volatile) |
10 setups and waveforms can be saved and recalled internally |
Storage mode |
USB 2.0 full speed host port (Support up to 4GB USB drive)
Image formats: .bmp (8-bit, 24-bit) and .png (24-bit)
Data format: .csv |
I/O |
USB 2.0 full-speed host, USB 2.0 full-speed client |
Printer compatibility |
PCL Inkjet, PCL Laser |
1. Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and within 23 ± 10 °C of last calibration temperature. |
Maximum input voltages |
Input CH1 and CH2 direct (1:1 probe) |
300 Vrms CAT III |
Input CH1 and CH2 (1:10 probe) |
600 Vrms 1 CAT III, 1000 Vrms 1 CAT II |
Input CH1 and CH2 (1:100 probe) |
600 Vrms 1 CAT II, 1000 Vrms 1 CAT II, 3540 Vrms 1 CAT I |
Meter input |
600 Vrms CAT III, 1000 Vrms CAT II |
Scope input |
300 Vrms CAT III |
Voltage ratings |
Vrms 50–60 Hz (AC sine wave), VDC (DC applications) |
General Specifications |
Power adapter |
Line voltage range: 50/60 Hz, 100 to 240 VAC, 1.6 A
Output voltage: 15 VDC, 4 A
Installation Category II |
Battery |
Li-Ion rechargeable battery pack, 10.8 V
Operating time: Up to 3 hours |
Operating environment |
Temperature |
0 to 50 °C (with battery only)
0 to 40 °C (with power adapter) |
Humidity |
0 to 80% RH (0 to 35 °C)
0 to 50% RH (35 to 40/50 °C)
Altitude up to 2000 m
Pollution degree 2 |
Storage compliance |
Temperature |
-20 to 70 °C |
Humidity |
0 to 80% RH
Altitude up to 15000 m |
Shock |
Tested to IEC 60068-2-27 |
Vibration |
Tested to IEC 60068-2-6, IEC 60068-2-64 |
Safety compliance |
IEC 61010-1:2001/EN 61010-1:2001
Canada: CAN/CSA-C22.2 No. 61010-1-04
USA: ANSI/UL 61010-1:2004 |
EMC compliance |
IEC 61326-1:2005/EN 61326-1:2006
Australia/New Zealand: AS/NZS CISPR 11:2004
Canada: ICES/NMB-001:ISSUE 4, June 2006 |
IP rating |
IP 41 ingress protection according to IEC 60529 |
Dimensions (W × H × D) |
183 x 270 x 65 mm |
Weight |
< 2.5 kg |
Warranty |
3 years for main unit