SCOPE SPECIFICATIONS [1] |
Vertical System: Scope Channels |
Bandwidth (–3 dB) |
U1602B: DC to 20 MHz
U1604B: DC to 40 MHz |
DC vertical gain accuracy |
5 mV/div to 20 mV/div: ± 5% full scale
50 mV/div to 100 V/div: ± 3% full scale |
Scope Channel Triggering |
Trigger sensitivity |
DC to 5 MHz: 0.8 divisions
U1602B: 5 MHz to 20 MHz – 1 division
U1604B: 5 MHz to 40 MHz – 1 division |
SCOPE CHARACTERISTICS [2] |
Acquisition: Scope Channels |
Maximum sample rate |
100 MSa/s per channel (50 s/div to 250 ns/div) 3
200 MSa/s single channel and interleaved (125 ns/div) 4 |
Equivalent sample rate |
U1604B: 2.5 GSa/s (125 ns/div to 10 ns/div) |
Vertical resolution |
8 bits |
Maximum recording length: |
U1602B/U1604B: Up to 125 000 points, viewable on screen with zoom function |
Peak detection |
5 ns |
Average |
Selectable in average number of 2, 4, 8, 16, 32, 64, 128, 256 |
Vertical System: Scope Channels |
Analog channels |
Channel 1 and Channel 2 simultaneous acquisition |
Bandwidth (–3 dB) |
U1602B: DC to 20 MHz
U1604B: DC to 40 MHz |
AC coupled |
< 10 Hz without probe
< 1 Hz with 10 MΩ 10:1 probe |
Rise time |
U1602B: < 17.5 ns
U1604B: < 8.8 ns |
Single shot bandwidth |
U1602B: 20 MHz
U1604B: 40 MHz |
Vertical sensitivity |
5 mV/div to 100 V/div (1:1 scope probe)
50 mV/div to 1 kV/div (10:1 scope probe)
500 mV/div to 10 kV/div (100:1 scope probe) |
Maximum input |
CAT III 300 Vrms (up to 400 Hz) from terminal to ground |
Offset/Dynamic range |
± 5 div |
Input impedance |
1 MΩ I I < 20 pF |
Coupling |
AC, DC, GND |
[1] All specifications are warranted. Specifications are valid after a 30-minute warm-up period and within a range of ±10 ºC from
firmware calibration temperature.
[2] All characteristics are typical performance values and are not warranted. Characteristics are valid after a 30-minute warm-up
period and within a range of ±10 ºC from firmware calibration temperature.
[3] Maximum sampling rates are shown here. Sampling rate will vary according to selected time base. For more information, please refer
to User’s and Service Guide.
[4] 200 MS/s sampling rate is only available at 125 ns/div timebase
[5] Number of points displayed will vary according to time base selected. For more information, refer to User’s and Service Guide |
Probes |
U1560-60001: 1:1 passive probe
U1561-60001: 10:1 passive probe
U1562-60001: 100:1 passive probe |
Probe attenuation factors |
1x, 10x, 100x |
Coupling |
3 Vp-p, ~ 1 kHz |
Maximum probe input |
1x CAT III 300 VAC
10 x, 100x CAT III 600 VAC |
Noise peak-to-peak |
3% of full scale or 5 mV, whichever is greater |
DC vertical offset accuracy |
± 0.1% div ± 2 mV ± 0.5% offset value |
Single cursor accuracy |
4% full scale |
Dual cursor accuracy |
4% full scale |
Horizontal System |
Range |
U1602B: 50 ns to 50 s/div
U1604B: 10 ns to 50 s/div |
Resolution |
U1602B: 2 ns
U1604B: 400 ps |
Reference position |
Left, center, right |
Delay range (pre-trigger) |
15 divisions |
Delay range (post-trigger) |
1000 divisions |
Analog Dt accuracy |
± 3% reading ± 0.4% screen |
Modes |
Main, XY, Roll |
RMS Jitter |
5% of horizontal scale or 5 ns, whichever is higher |
Trigger System |
Source |
Channel 1 and Channel 2 |
Modes |
Auto, normal, single |
Selections |
Edge |
Edge, pulse width, pattern, video
Trigger on a rising or falling edge of any source |
|
Pattern |
Trigger at the beginning of a pattern of high, low levels and rising or falling
edge established conditions of AND, OR, NOR and NAND between the
channels. |
|
Pulse Width |
200 ns to 10 s. Trigger when a positive or negative pulse width of any source
larger than, less than, equal to or not equal to duration. |
|
Video |
Video trigger sensitivity: 0.7 division trigger level.
Available to both Channel 1 and Channel 2.
Analog progressive and interlaced video standards including NTSC,
PAL and SECAM.
Positive or negative sync pulse polarity.
Modes – all fields, even fields, odd fields or line 5 – 263 within a field. |
Range |
± 4 divisions from center screen |
Level accuracy |
± 0.5 divisions |
Trigger sensitivity |
DC to 5 MHz: 0.8 divisions
U1602B: 5 MHz to 20 MHz – 1 division
U1604B: 5 MHz to 40 MHz – 1 division |
Coupling |
DC, AC (< 1 Hz), HF reject (> 50 kHz), LF reject (< 30 kHz), Noise reject |
Measurement System |
Autoscale |
Finds and displays all active scope channels, sets edge trigger mode on
highest numbered channel, sets vertical sensitivity on scope channel.
Requires voltage > 20 mVp-p, 0.5% duty cycle and frequency > 100 Hz. |
Automatic measurement |
Measurements continuously updated. |
Voltage |
Peak-to-peak, maximum, minimum, amplitude, top, base, +overshoot,
–overshoot, preshoot, RMS, mean and one cycle mean. |
Time |
Frequency, period, +width, –width, and +duty cycle and –duty cycle on
any channel.
Rise time, fall time, delay and phase shift. |
Cursors |
Manually place readout of horizontal (X, DX) and vertical (Y, DY). |
Waveform math |
CH1 + CH2, CH1 – CH2, CH2 – CH1 |
FFT [1] |
Window |
Rectangular, Hanning, Hamming, Blackman-Harris |
Amplitude display |
Selectable in amplitude displays of 1 dB, 2 dB, 5 dB, 10 dB |
Digital Multimeter Specifications [1] ± (% reading + % range) |
Function |
Range |
Frequency, Test Current or Burden Voltage |
1 year Tcal ± 5º C |
DC Voltage |
600.0 mV
6.000 V
60.00 V
600.0 V |
|
0.3 + 0.08
0.3 + 0.08
0.3 + 0.08
0.3 + 0.08 |
AC Voltage |
600.0 mV – 600.0 V |
50 Hz – 1 kHz
1 kHz – 30 kHz |
1.0 + 0.2
3.0 + 0.2 |
AC + DC Voltage |
6.000 V – 600.0 V |
50 Hz – 1 kHz
1 kHz – 30 kHz |
1.0 + 0.2
3.0 + 0.2 |
Resistance |
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
60.00 MΩ |
|
0.5 + 0.2
0.5 + 0.2
0.5 + 0.2
0.5 + 0.2
0.5 + 0.2
1.0 + 0.2 |
Capacitance |
60.00 nF
600.0 nF
6.000 µF
60.00 µF
300.0 µF |
|
2.0 + 0.2
2.0 + 0.2
2.0 + 0.2
2.0 + 0.2
2.0 + 0.2 |
Diode |
1.000 V |
0.5 mA |
2.0 + 0.08 |
[1] For temperatures between 0 ºC to 18 ºC and 28 ºC to 50 ºC, add 0.1% of reading + 0.02% of range for every degree Celsius. |
Digital Multimeter Specifications [1] ± (% reading + % range) |
Function |
Range |
Frequency |
1 year Tcal ± 5º C |
Temperature [2] , ºC |
600.0 ºC
6000 ºC |
|
0.3 + 0.08
6000 ºC 0.3 + 0.08 |
Temperature [2] , ºF |
6000 ºCF
6000 ºF |
|
0.3 + 0.08
0.3 + 0.08 |
AC Current [3] |
60.00 A
600.0 A |
50 Hz – 1 kHz
50 Hz – 1 kHz |
1.0 + 0.2
1.0 + 0.2 |
Measurement Characteristics |
Full scale reading |
6,000-count |
DC voltage, True RMS AC voltage |
Maximum input voltage, 600 Vrms CAT II, 300 Vrms CAT III
DC coupled input coupling |
Continuity |
Beeper < 60 W in 600 W range |
Data Logger |
Source |
Digital multimeter measurements |
Range |
10 divisions |
Record size |
Up to 8800 data points (with option 001) |
Time span |
Auto range 150 seconds to 20 days |
Time reference |
Time from start |
Record method |
Selectable minimum, maximum and average |
Display System |
Display |
4.5-inch diagonal color CSTN LCD |
Resolution |
320 x 240 pixels |
Control |
Contrast control, infinite persistance on/off |
Built-in help system |
Functional help displayed by pressing help button |
Real-time clock |
Time and date (user-adjustable) |
Storage |
Save/Recall (non-volatile) |
Up to 10 setups and traces |
[2] U1586B temperature module is needed to make measurements and is available as option. [3] U1583B AC current clamp needed to make measurements and is available as option. |
GENERAL CHARACTERISTICS |
Power Adapter |
Line voltage range 50/60 Hz, 100 – 240 VAC
Output voltage 12 VDC |
Battery |
Ni-MH rechargeable battery pack 7.2 V, 4500 mAH
Operating time: 4 hours
Charging time: 5 hours, measurement unit off
Allow ambient temperature during charging: 10 ºC to 40 ºC |
Operating Environment |
Temperature |
Operating full accuracy
Non-operating |
0 ºC to 50 ºC
–20 ºC to 70 ºC |
Humidity |
Operating full accuracy |
to 80% RH at 40 ºC |
Altitude |
Operating full accuracy
Non-operating |
Up to 2000 m
15000 m (50000 ft) |
ESD tolerance |
± 4 kV |
Safety Compliance |
IEC 61010-1: 2001/EN61010-1: 2001
CSA C22.2 No. 61010-1: 2004
Pollution degree 2
This instrument is rated for indoor use only. |
Dimensions (HxWxD) |
24.1 cm height x 13.8 cm width x 6.6 cm depth |
Weight |
1.5 kg |
I/O |
USB 2.0 Full Speed Client - For PC connectivity (Standard Option)
USB 2.0 Full Speed Host - For external USB storage (Option 001)
Firmware upgrade through USB. |
Warranty |
3 years for main unit
3 months for standard shipped accessories unless otherwise stated. |