Function |
90 Day
23°C ± 5°C |
1 Year
23°C ± 5°C |
Temperature Coefficient
0°C – 18°C
28°C – 55°C |
dc Voltage |
0.0040 + 0.0035
0.0030 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0006
0.0035 + 0.0010 |
0.0050 + 0.0035
0.0040 + 0.0007
0.0035 + 0.0005
0.0045 + 0.0006
0.0045 + 0.0010 |
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001 |
True rms ac Voltage4 |
1.00 + 0.04
0.35 + 0.04
0.05 + 0.04
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50 |
1.00 + 0.04
0.35 + 0.04
0.06 + 0.04
0.12 + 0.04
0.60 + 0.08
4.00 + 0.50 |
0.100 + 0.004
0.035 + 0.004
0.005 + 0.004
0.011 + 0.005
0.060 + 0.008
0.20 + 0.02 |
1.00 + 0.03
0.35 + 0.03
0.05 + 0.03
0.11 + 0.05
0.60 + 0.08
4.00 + 0.50 |
1.00 + 0.03
0.35 + 0.03
0.06 + 0.03
0.12 + 0.04
0.60 + 0.08
4.00 + 0.50 |
0.100 + 0.003
0.035 + 0.003
0.005 + 0.003
0.011 + 0.005
0.060 + 0.008
0.20 + 0.02 |
Resistance7 |
0.008 + 0.004
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.008 + 0.001
0.020 + 0.001
0.800 + 0.010 |
0.010 + 0.004
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.010 + 0.001
0.040 + 0.001
0.800 + 0.010 |
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002 |
dc Current |
0.030 + 0.020
0.030 + 0.005
0.080 + 0.010
0.120 + 0.020 |
0.050 + 0.020
0.050 + 0.005
0.100 + 0.010
0.120 + 0.020 |
0.002 + 0.0020
0.002 + 0.0005
0.005 + 0.0010
0.005 + 0.0020 |
True rms ac Current4 |
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04 |
1.00 + 0.04
0.30 + 0.04
0.10 + 0.04 |
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006 |
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06 |
1.10 + 0.06
0.35 + 0.06
0.15 + 0.06 |
0.100 + 0.006
0.035 + 0.006
0.015 + 0.006 |
Frequency or Period8 |
0.10
0.05
0.03
0.01 |
0.10
0.05
0.03
0.01 |
0.005
0.005
0.001
0.001 |
Continuity |
0.008 + 0.020 |
0.010 + 0.020 |
0.001 + 0.002 |
Diode Test |
0.008 + 0.020 |
0.010 + 0.020 |
0.001 + 0.002 |
- Specifications are for 1hr warm-up and 61/2 digits, Slow ac filter.
- Relative to calibration standards.
- 20% over range on all ranges except 1000Vdc and 750Vac ranges.
- For sinewave input > 5% of range. For inputs from 1% to 5% of range and < 50kHz, add 0.1% of range additional error.
- 750V range limited to 100 kHz or 8 x107 Volt-Hz.
- Typically 30% of reading error at 1MHz.
- Specifications are for 4- wire ohms function or 2-wire ohms using Math Null. Without Math Null, add 0.2 Ω additional error in 2-wire ohms function.
- Input > 100 mV. For 10 mV inputs multiply % of reading error x10.
|