Specification |
Accuracy |
ADT672-02: 0.02% of full scale |
ADT672-05: 0.05% of full scale |
>20,000 (1,400 bar): 0.1% FS |
Gauge Types |
Gauge pressure, compound pressure, absolute pressure, differential pressure |
Display |
Description: Dual-line 6 full digit FSTN LCD with LED Backlight |
Display rate: 3.5 readings per second (Default setting). |
Numeral display height: 16.5mm (0.65") |
Pressure Units |
Pa, kPa, MPa, psi, bar, mbar, kgf/cm2, inH2O@4°C mmH2O@4°C, inHg@0°C, mmHg@0°C |
Environmental |
Compensated Temperature: 14°F to 122°F (-10°C to 50°C) |
Operating Temperature: 14°F to 122°F (-10°C to 50°C) |
Storage Temperature: -4°F to 158°F (-20°C to 70°C) |
Humidity:<95% |
Pressure Port |
≤15,000 psi: 1/4NPT male, 1/2NPT male, 1/4BSP male, 1/2BSP male, M20×1.5 male |
>15,000 psi: 1/4HP female or 1/4HP male *1/4HP female: Autoclave F-250-C, 9/16" - 18 UNF-2B *1/4HP male: Autoclave M-250-C, 9/16" - 18 UNF-2A |
Differential Pressure: 0.236 inch (Ø6 mm) test hose |
Other connections available per request |
Over Pressure Warning |
1.2 |
Electrical Connection |
0.156 inch (Ø4mm) sockets |
Electrical Measurement Accuracy |
Voltage DC: ±30.0000V, ± (0.01%RD + 1.5 mV) |
Current DC: ±30.0000mA, ± (0.01%RD + 1.5 μA) |
DC 24V: 24V±0.5V, MAX:50mA, Protect at: 120mA |
Switch[1]: Status OPEN/CLOSED |
Power |
Battery: Rechargeable Li-ion polymer battery |
Li-Battery working time: 40 hours |
Recharge time: 4 hours |
External power: 110V/220V power adapter (DC10V) |
Enclosure |
Case material: Aluminum alloy |
Wetted parts: 316L SS |
Dimension: Ø120mm X 46mm depth X 184mm height |
Weight: 0.7kg |
Data Logging |
Storage capacity: 30 files, 40 records per file |
Mode: manual and automatic |
Hourly-record: record the data every hour |
Interval-record: set by user |
Compliance |
CE Marked |
Communication |
RS232 (DB9/F, environmentally sealed) |
Baud rate: 1200, 2400, 4800, 9600 |
Data length: 8 bits |
Stop bit: 2 bits |
Address: from 1 to 112 |
Warranty |
1 year |
[1] 1V~12V if switch has detective voltage |