|
OX 5022 |
OX 5042 |
Type of display |
3.5” colour TFT LCD screen – Resolution 320x240 –LED backlighting |
Display mode |
2,500 real acquisition points on screen |
Display of curves on screen |
2 curves + 2 references + memory trace or mathematical calculation |
Commands |
Direct adjustments on front panel & on-screen menus via browser (principal & secondary without “hidden menus”) |
Integrated interactive help function |
11 languages: French, English, German, Spanish, Italian, Swedish, Rumanian, Russian, Finnish, Polish, Dutch |
Oscilloscope Mode |
Bandwidth |
20 MHz |
40 MHz |
Bandwidth limiter |
1.5 MHz, 5 kHz |
Number of channels |
2 totally-isolated channels |
Input impedance |
1 MΩ ±0.5%, approx. 17 pF |
Maximum input voltage |
600 V CAT III – Derating -20dB per decade from 100 kHz |
Vertical sensitivity |
5 mV to 200 V/div |
Sweep speed |
25 ns/div to 200 s/div –Roll Mode from 100 ms to 200 s/div |
Horizontal zoom |
Zoom factor: x1, x2, x5 |
Mode |
Automatic, triggered, one-shot & triggered Roll |
Type |
Edge, pulse width (20 ns – 20 s) |
Coupling |
AC or DC (depending on the coupling of the triggering channel) - HF, LF or noise rejection |
Sensitivity |
≤ 1.2 divisions p-p up to 20 MHz |
≤ 1.2 divisions p-p up to 40 MHz |
Maximum sampling rate |
2 GS/s in ETS mode – 50 MS/s in one-shot mode on each channel |
Vertical resolution |
9 bits |
Memory depth |
2,500 points per channel |
User storage |
2 MB for storing files: trace (.trc), text, (.txt), con?guration (.cfg), image files (.bmp) |
GLITCH mode |
Duration ≥ 20 ns – 1,250 Min/Max pairs |
Display modes |
Envelope, Averaging (factors 2 to 64) and XY (vector) |
MATH functions |
Channel inversion, addition, subtraction, multiplication and division (adjustable scaling) |
Cursor measurements |
2 cursors: V, T, dV, dt simultaneously –4-digit display resolution |
Automatic measurements |
18 time or level measurements and phase measurement |
General specifications |
2 channels, 8,000-count display + min/max bargraph – Graphic recording of 2,700 measurements (5 min to 1 month) |
Operating modes |
Absolute or relative display (absolute, deviation, ref, ref%) – Monitoring (instantaneous, Min, Max, Avg) |
AC, DC and AC+DC voltages |
Ranges from 600 mV to 600 VRMS, 800 mV to 800 VDC – accuracy for VDC 1%reading+20D –50 kHz bandwidth |
Resistance |
Range from 80 O to 32 MO - accuracy 2%R + 10D –10 ms quick continuity test |
Capacitance |
Ranges from 5 nF to 5 mF – basic accuracy 2%reading+10D |
Other measurements |
Frequency, rotation speed, 3.3 V diode test, temperature measurement (with K thermocouple or infrared probe) |
Measurements |
Single-phase and balanced three-phase active power values (with or without neutral), simultaneous display of current |
Multi-channel analysis |
2 channels, 31 orders, fundamental frequency from 40 to 450 Hz |
Simultaneous measurements |
Total V RMS , THD and selected order (%fundamental, phase, frequency, V RMS ) |
Screenshots |
Up to 100 file in standard “.bmp” format, viewable on the instrument |
PC communication |
Isolated optical USB interface –“SX-Metro” PC application software available as an option (version CK) |
Power supply |
6 LR6 or 6 AA NiMh batteries – Battery life up to 8 hrs 30 min – Universal mains adapter isolated from the channels – Quick charging in 2 hrs 30 min |
Safety / EMC |
Safety according to IEC61010-1 Ed3 – 600 V CAT III – EMC according to EN61000-3, 2001 & EN61326-1, 2006 |
Mechanical specifications |
214x110x57mm – 1.2 kg with batteries – moulded elastomer casing, IP54 protection |
Warranty |
3 years |