AEMC 2116.72 Specifications |
Electrical Specifications |
Reserve Power Function |
Yes. Convert VAC + DC (True RMS) in power (from 100mVA to 100kVA), with a resistance reference adjustable from 1 to 9999W. |
AC Current |
RANGE |
500µA / 5mA / 50mA / 500mA / 10A |
SCALE |
5% to 100% of range |
FREQUENCY RANGE |
DC (AC+DC only) or 40Hz to 5kHz |
ACCURACY |
0.75% Reading + 30D / 0.6% Reading + 30D / 0.7% Reading + 30D / 1% Reading + 30D up to 2kHz |
PROTECTION |
600VRMS |
FUSES |
F1 + F2 |
RESOLUTION |
10nA / 100nA / 1µA / 10µA / 1mA |
DC Voltage |
SELECTOR SWITCH POSITION |
mV / VDC |
RANGE |
500mV / 5V / 50V / 500V / 1000V |
ACCURACY |
0.05% Reading + 2D |
INPUT IMPEDANCE |
10MW/1GW (refer to § 3.4) / 11MW / 10MW |
PROTECTION |
±1100 VPK (at 23°C ±2°C) / ±1100 VPK |
RESOLUTION |
10µV / 100µV / 1mV / 10mV / 100mV |
DC Current |
SELECTOR SWITCH POSITION |
µA / mA / 10A |
RANGE |
500µA / 5mA / 50mA / 500mA / 10A |
ACCURACY |
0.2% Reading + 5D / 0.2% Reading + 2D / 0.05% Reading + 2D 0.2% Reading + 2D / 0.5% Reading + 5D |
MAX VOLTAGE DROP |
700mV / 1.5V / 500mV |
PROTECTION |
600VRMS |
FUSES |
F1 + F2 / F2 |
RESOLUTION |
10nA / 100nA / 1µA / 10µA / 1mA |
Resistive Power Function |
MEASURED FUNCTION |
(measured voltage)2/Rref |
RESOLUTION |
100µW |
ACCURACY |
2 x VAC accuracy (%) |
MEASUREMENT MAX VOLTAGE |
750VAC+DC |
PROTECTION |
±100VPK |
DISPLAY UNIT |
VA |
Diode Test |
MEASURABLE VOLTAGES |
0 to 2V |
MEASUREMENT CURRENT |
1mA ±20% |
RESOLUTION |
1mV |
PROTECTION |
600VRMS can be reset automatically |
Duty Cycle |
RESOLUTION |
0.01% |
MINIMUM DURATION FOR 0 OR T-0 |
2µs |
MAXIMUM DURATION FOR T |
0.8s |
MINIMUM DURATION FOR T |
100µs |
ABSOLUTE ERROR AS A % |
3.10-4/T |
SENSITIVITY |
see sensitivity in frequency measurement mode |
ABSOLUTE ADDITIONAL ERROR |
0.1 x C/P, with C: V or A range (C=5000V for 1000VDC or 750VAC ranges and C=50A for 10A range, and with P: slope in V/s or in A/s |
Noise |
PRESENCE OF SPIKES |
>80V, in the 1kHz - 100kHz band, by a 3 bar l l l display next to the symbol |
PRESENCE OF NOISE LEVEL |
>25V in the 1kHz to 1000kHz band, by a bar l l l display next to the symbol |
THE RMS VALUE OF THE MAINS VOLTAGE |
50 or 60Hz in the 500VAC range, accurate to 1% |
Pulse Count |
MINIMUM PULSE DURATION |
2µs |
COUNT |
up to 99999 |
LIMIT OF TRIGGERING |
3% of range except in 750VAC-1000VDC range (limit: 150V) and 10A (limit: 1.5A). |
COMMENT |
Reset the count to zero with the REL key |
Pulse Width |
RESOLUTION |
10µs |
MINIMUM PULSE WIDTH |
20µs |
ACCURACY |
0.01% ±10µs |
MAXIMUM DURATION OF A PERIOD |
12.5s |
LIMIT OF TRIGGERING |
3% of range except in 750VAC-1000VDC range (limit: 150V) and 10A (limit: 1.5A). This limit is positive in ( ) and negative in ( ). |
COMMENT |
Additional error on measurement due to the zero-crossing slope: refer to duty cycle measurement |
Measurement Rate |
Digital Display |
2 measurements/s |
Bargraph |
20 measurements/s |
General Specifications |
Power Supply |
9V alkaline battery (6LF22) |
Battery Life |
The typical life of 500 hours in VDC mode |
Mechanical Dimensions |
7.4 x 3.2 x 1.6" (189 X 82 X 40 mm) |
Packing Dimensions |
9.1 x 6.1 x 2.6" (230 X 155 X 65mm) |
Weight |
.9 lbs (400g) |
Packaging Weight |
500g |
Environmental Specification |
Environment |
Indoor use |
Altitude |
<2000m |
Reference Temperature |
18°C to 28°C |
Humidity |
RANGE |
0 to 40°C (70% max for 5MW/50MW) / 40°C to 50°C / above 50°C |
RELATIVE HUMIDITY |
0 to 80% / 0 to 70% / 60% |
Rated Range of Use |
0°C to 50°C |
Limit Range of Operation |
-10°C to 60°C |
Storage Temperature Range |
-40°C to 70°C |
Temperature Coefficient |
max 0.1 x accuracy /°K |
Casing and Circuit |
self-extinguishing materials |
Operating Quality |
IEC 359 |
Safety Reference Specifications |
Electrical Safety (per brochure) |
IEC 1010-1 |
Electromagnetic Compatibility |
NF EN 61326-1 |
Display |
Liquid crystal display comprising: a 50,000 count display + sign (digits 14mm high), a 34 bar analog bar graph display, appropriate units for each type of measurement |